3.10 Medical physics (A-level only) — coverage pack

19 specification leaves · notes, questions, answers and worked methods

3.10.1.1 · Physics of vision

  • The cornea supplies most of the eye's refraction; the lens changes power during accommodation so that a real image forms on the retina.
  • Cone cells work best in bright light, provide colour vision and give high spatial resolution because their nerve pathways show little convergence.
  • Rod cells are more sensitive at low light levels but do not distinguish colour; many rods share a nerve pathway, reducing spatial resolution.
  • The eye's spectral response depends on wavelength and on which receptors are active. A common error is to describe rods as producing detailed colour vision.

Tier 1 · Easy

  1. 1. State the type of retinal receptor that is mainly responsible for vision at very low light intensity and give one property of the resulting image.[2 marks]

    Answer

    • Rod cells; the image has no colour or has low spatial resolution.

    Method: Low light levels stimulate the more sensitive rod cells. Rods do not provide colour vision, and the convergence of several rods onto one nerve pathway gives low spatial resolution; either image property earns the second marking point.

Tier 2 · Standard

  1. 1. Explain why two faint points of light that fall on nearby rod cells may be seen as one point even though both points are detected.[3 marks]

    Answer

    • Several rod cells can share one nerve pathway, so signals from the two illuminated rods may be combined and the brain cannot distinguish their positions; sensitivity is increased but spatial resolution is reduced.

    Method: Identify convergence: several rods connect to a common sensory pathway. Their signals are combined, which helps a small light stimulus reach the detection threshold. Because the same pathway does not identify which individual rod responded, the two source positions cannot be resolved.

Tier 3 · Hard

  1. 1. A coloured grid is viewed first in bright light and then after the illumination is greatly reduced. Explain the changes expected in colour, detail and sensitivity, referring to the retinal receptors and their nerve connections.[5 marks]

    Answer

    • In bright light cones respond, so the grid is seen in colour and with high spatial resolution because cone pathways show little convergence. In dim light rods dominate, so colour is lost and fine lines are less easily resolved because many rods share a pathway, although this convergence makes the eye more sensitive to weak light.

    Method: For bright light, award the linked chain cones, colour vision, and high resolution from separate or weakly convergent pathways. For dim light, award rods, no colour, and reduced resolution from convergence. Complete the comparison by linking that convergence to greater sensitivity because signals from several rods can combine.

3.10.1.2 · Defects of vision and their correction using lenses

  • Lens power is P=1/fP=1/f with ff in metres and PP in dioptres: converging lenses have positive power and diverging lenses have negative power.
  • Using the real-is-positive convention, 1f=1u+1v\frac{1}{f}=\frac{1}{u}+\frac{1}{v}; a virtual object or image distance is negative, and m=v/um=v/u.
  • Myopia focuses distant objects in front of the retina and is corrected by a diverging lens; hypermetropia needs a converging lens to view near objects clearly.
  • Astigmatism is corrected with a cylindrical component whose prescription includes its power and an axis between 00^\circ and 180180^\circ.
  • Keep all distances in metres before finding power. A common error is to use the magnitude of a virtual image distance as positive.

Tier 1 · Easy

  1. 1. State the type of spectacle lens used to correct myopia and state the sign of its power.[2 marks]

    Answer

    • A diverging (concave) lens with negative power.

    Method: A myopic eye is too powerful, so parallel rays would focus before the retina. A diverging lens reduces the convergence before the rays enter the eye; its focal length and power are negative.

Tier 2 · Standard

  1. 1. A converging lens has power +5.0D+5.0\,\text{D}. An object is 0.60m0.60\,\text{m} from the lens. Using the real-is-positive convention, calculate the image distance and the magnitude of the magnification.[3 marks]

    Answer

    • v=+0.30mv=+0.30\,\text{m} and m=0.50|m|=0.50

    Method: The focal length is f=1/P=1/5.0=0.20mf=1/P=1/5.0=0.20\,\text{m}. Hence 1/v=1/f1/u=5.01/0.60=3.33m11/v=1/f-1/u=5.0-1/0.60=3.33\,\text{m}^{-1}, so v=+0.30mv=+0.30\,\text{m}. The magnification magnitude is m=v/u=0.30/0.60=0.50|m|=|v/u|=0.30/0.60=0.50.

Tier 3 · Hard

  1. 1. A hypermetropic eye has a near point 0.80m0.80\,\text{m} from a spectacle lens. Determine the lens power needed so that an object 0.25m0.25\,\text{m} from the lens forms a virtual image at the near point. Use the real-is-positive convention.[5 marks]

    Answer

    • +2.8D+2.8\,\text{D}

    Method: The object is real, so u=+0.25mu=+0.25\,\text{m}. The required image is virtual, so v=0.80mv=-0.80\,\text{m}. Then 1/f=1/u+1/v=1/0.25+1/(0.80)=4.001.25=2.75m11/f=1/u+1/v=1/0.25+1/(-0.80)=4.00-1.25=2.75\,\text{m}^{-1}. Therefore P=1/f=+2.75DP=1/f=+2.75\,\text{D}, which is +2.8D+2.8\,\text{D} to 22 significant figures. The positive sign confirms that a converging lens is required.

3.10.2.1 · Ear as a sound detection system

  • The pinna and ear canal direct sound to the tympanic membrane, whose vibrations move the ossicles in the middle ear.
  • The ossicles act as a lever system, and the smaller area of the oval window than the tympanic membrane produces a larger pressure in the cochlear fluid.
  • Pressure waves in the cochlea move regions of the basilar membrane; hair cells convert this mechanical motion into electrical signals in the auditory nerve.
  • When explaining transmission, follow the complete chain from pressure variation in air to mechanical vibration, fluid motion and an electrical nerve signal.

Tier 1 · Easy

  1. 1. Name the membrane that vibrates when sound reaches the end of the ear canal and name the three-bone system moved by it.[2 marks]

    Answer

    • The tympanic membrane (eardrum) and the ossicles.

    Method: Sound pressure variations act on the tympanic membrane. Its vibration is transmitted through the malleus, incus and stapes, collectively called the ossicles.

Tier 2 · Standard

  1. 1. Explain two features of the middle ear that increase the pressure delivered to the fluid in the cochlea.[3 marks]

    Answer

    • The ossicles provide a lever force gain, and this force acts on the oval window, whose area is much smaller than the tympanic membrane; since pressure is force divided by area, the pressure increases.

    Method: First identify the lever action of the ossicles and link it to an increased force. Then compare areas: the oval window is smaller than the tympanic membrane. Using p=F/Ap=F/A, a larger force acting over a smaller area gives a greater pressure in the inner-ear fluid.

Tier 3 · Hard

  1. 1. A sound produces pressure amplitude 0.012Pa0.012\,\text{Pa} on a tympanic membrane of area 5.5×105m25.5\times10^{-5}\,\text{m}^2. The ossicles increase the force by a factor of 1.41.4 and act on an oval window of area 3.2×106m23.2\times10^{-6}\,\text{m}^2. Calculate the pressure amplitude delivered to the inner-ear fluid.[5 marks]

    Answer

    • 0.29Pa0.29\,\text{Pa}

    Method: The force on the tympanic membrane is F1=p1A1=0.012(5.5×105)=6.6×107NF_1=p_1A_1=0.012(5.5\times10^{-5})=6.6\times10^{-7}\,\text{N}. The ossicle output force is F2=1.4F1=9.24×107NF_2=1.4F_1=9.24\times10^{-7}\,\text{N}. Therefore p2=F2/A2=(9.24×107)/(3.2×106)=0.28875Pa=0.29Pap_2=F_2/A_2=(9.24\times10^{-7})/(3.2\times10^{-6})=0.28875\,\text{Pa}=0.29\,\text{Pa} to 22 significant figures.

3.10.2.2 · Sensitivity and frequency response

  • Sound intensity is power transferred per unit area, measured in W m2\text{W m}^{-2}; for an isotropic point source, I=P/(4πr2)I=P/(4\pi r^2).
  • Intensity level is L=10log10(I/I0)L=10\log_{10}(I/I_0) in decibels, where I0=1.0×1012W m2I_0=1.0\times10^{-12}\,\text{W m}^{-2}.
  • Equal-loudness curves show the intensity level needed at each frequency for the same perceived loudness; normal hearing is most sensitive around a few kilohertz.
  • The logarithmic decibel scale reflects human perception: a change in intensity level is found from ΔL=10log10(I2/I1)\Delta L=10\log_{10}(I_2/I_1), not from an intensity ratio in decibels.
  • The dBA scale applies frequency weighting to approximate the response of human hearing; do not confuse dBA weighting with an unweighted dB intensity level.

Tier 1 · Easy

  1. 1. Calculate the intensity level of a sound with intensity 1.0×108W m21.0\times10^{-8}\,\text{W m}^{-2}. Use I0=1.0×1012W m2I_0=1.0\times10^{-12}\,\text{W m}^{-2}.[2 marks]

    Answer

    • 40dB40\,\text{dB}

    Method: Use L=10log10(I/I0)L=10\log_{10}(I/I_0). Thus L=10log10((1.0×108)/(1.0×1012))=10log10(104)=40dBL=10\log_{10}((1.0\times10^{-8})/(1.0\times10^{-12}))=10\log_{10}(10^4)=40\,\text{dB}.

Tier 2 · Standard

  1. 1. Two sounds have intensity levels 76dB76\,\text{dB} and 58dB58\,\text{dB}. Determine the ratio of the larger intensity to the smaller intensity.[3 marks]

    Answer

    • 6363

    Method: The level difference is ΔL=7658=18dB\Delta L=76-58=18\,\text{dB}. From ΔL=10log10(I2/I1)\Delta L=10\log_{10}(I_2/I_1), the ratio is I2/I1=1018/10=101.8=63.1I_2/I_1=10^{18/10}=10^{1.8}=63.1, so the larger intensity is about 6363 times the smaller.

Tier 3 · Hard

  1. 1. An isotropic sound source has power 0.80W0.80\,\text{W}. Determine the distance at which its intensity level is 85dB85\,\text{dB}. Then explain why a 12kHz12\,\text{kHz} tone at this level may sound quieter than a 3kHz3\,\text{kHz} tone at the same level. Use I0=1.0×1012W m2I_0=1.0\times10^{-12}\,\text{W m}^{-2}.[5 marks]

    Answer

    • 14m14\,\text{m}; normal hearing is less sensitive at 12kHz12\,\text{kHz} than near 3kHz3\,\text{kHz}, so the higher-frequency tone has a greater equal-loudness threshold.

    Method: First find the intensity: I=I010L/10=(1.0×1012)108.5=3.16×104W m2I=I_0 10^{L/10}=(1.0\times10^{-12})10^{8.5}=3.16\times10^{-4}\,\text{W m}^{-2}. For an isotropic source, I=P/(4πr2)I=P/(4\pi r^2), so r=P/(4πI)=0.80/(4π(3.16×104))=14.2mr=\sqrt{P/(4\pi I)}=\sqrt{0.80/(4\pi(3.16\times10^{-4}))}=14.2\,\text{m}, giving 14m14\,\text{m} to 22 significant figures. Equal-loudness curves have their minimum near a few kilohertz; at 12kHz12\,\text{kHz} a larger level is required for the same perceived loudness.

3.10.2.3 · Defects of hearing

  • Hearing loss raises the intensity level required for a sound to be heard, so the affected equal-loudness or threshold curve shifts upward over the damaged frequency range.
  • Age-related deterioration is typically greatest at high frequencies, whereas prolonged excessive noise exposure commonly produces the greatest loss near 4kHz4\,\text{kHz}.
  • A hearing loss in decibels is the vertical difference between normal and impaired threshold curves at the stated frequency.
  • The specification tests changes in hearing and equal-loudness curves, not the physiological mechanism of the damage; avoid unsupported descriptions of structures being damaged.

Tier 1 · Easy

  1. 1. State the frequency range in which age-related hearing deterioration is usually greatest.[1 mark]

    Answer

    • At high frequencies.

    Method: Age-related hearing loss does not raise the threshold equally at all frequencies. The characteristic change is a larger loss towards the high-frequency end of an equal-loudness or threshold graph.

Tier 2 · Standard

  1. 1. At 4.0kHz4.0\,\text{kHz}, a normal threshold is 8dB8\,\text{dB} and an impaired threshold is 38dB38\,\text{dB}. Calculate the hearing loss and the factor by which the threshold intensity has increased.[3 marks]

    Answer

    • 30dB30\,\text{dB} and a factor of 1.0×1031.0\times10^3

    Method: The hearing loss is the vertical threshold difference: 388=30dB38-8=30\,\text{dB}. Using ΔL=10log10(I2/I1)\Delta L=10\log_{10}(I_2/I_1) gives I2/I1=1030/10=103I_2/I_1=10^{30/10}=10^3. The impaired threshold intensity is therefore 10001000 times the normal value.

Tier 3 · Hard

  1. 1. Listener A has hearing losses of 6dB6\,\text{dB} at 0.5kHz0.5\,\text{kHz}, 15dB15\,\text{dB} at 4kHz4\,\text{kHz} and 44dB44\,\text{dB} at 10kHz10\,\text{kHz}. Listener B has losses of 9dB9\,\text{dB}, 48dB48\,\text{dB} and 17dB17\,\text{dB} at the same frequencies. Deduce the likely cause for each pattern and explain how each pattern changes an equal-loudness curve and perceived sound.[5 marks]

    Answer

    • A is consistent with age-related deterioration because the loss rises strongly at high frequency. B is consistent with excessive-noise exposure because its largest loss is near 4kHz4\,\text{kHz}. Each equal-loudness curve is shifted upward most in its affected range, so greater intensity is needed there and sounds at those frequencies are perceived as quieter.

    Method: Compare the frequency dependence rather than the total loss. A's largest value is at 10kHz10\,\text{kHz}, matching age-related high-frequency loss. B has a pronounced maximum at 4kHz4\,\text{kHz}, matching excessive-noise exposure. A positive hearing loss raises the required level on the equal-loudness curve; the larger the loss, the larger the upward displacement and the quieter an unchanged sound is perceived.

3.10.3.1 · Simple ECG machines and the normal ECG waveform

  • Skin electrodes measure small potential differences produced by the heart; conductive gel, clean skin and secure electrodes reduce contact resistance and movement artefacts.
  • A differential, high-gain, low-noise amplifier and shielded leads are used because the ECG signal is only of order millivolts and is vulnerable to electrical interference.
  • The P wave represents atrial depolarisation, the QRS complex ventricular depolarisation, and the T wave ventricular repolarisation.
  • Heart rate is found from successive equivalent points, normally R peaks: rate=60/T\text{rate}=60/T when the period TT is in seconds. Measure several cycles to reduce percentage uncertainty.
  • Do not describe the ECG as a direct trace of blood pressure or mechanical force; it records electrical potential difference.

Tier 1 · Easy

  1. 1. Identify the electrical events represented by the QRS complex and the T wave in a normal ECG.[2 marks]

    Answer

    • The QRS complex is ventricular depolarisation and the T wave is ventricular repolarisation.

    Method: Match each named feature to the electrical activity of the ventricles: rapid ventricular depolarisation produces QRS, and the later recovery or repolarisation produces T.

Tier 2 · Standard

  1. 1. Five successive R peaks occur at 0.42s0.42\,\text{s}, 1.20s1.20\,\text{s}, 1.98s1.98\,\text{s}, 2.76s2.76\,\text{s} and 3.54s3.54\,\text{s}. Determine the mean heart rate in beats per minute.[3 marks]

    Answer

    • 77beats min177\,\text{beats min}^{-1}

    Method: The five peaks span four complete cardiac periods. Their total time is 3.540.42=3.12s3.54-0.42=3.12\,\text{s}, so T=3.12/4=0.780sT=3.12/4=0.780\,\text{s}. The rate is 60/T=60/0.780=76.9beats min160/T=60/0.780=76.9\,\text{beats min}^{-1}, which rounds to 77beats min177\,\text{beats min}^{-1}.

Tier 3 · Hard

  1. 1. An ECG trace contains slow baseline changes when the patient moves and a regular interference signal from nearby mains equipment. Explain how electrode attachment, the leads and the amplifier should be arranged to obtain a clearer normal waveform.[5 marks]

    Answer

    • Clean the skin and use low-resistance conductive gel, then secure non-reactive electrodes so movement does not change the contact. Use shielded leads kept away from mains sources and a differential low-noise amplifier. The amplifier needs high gain because the cardiac potential differences are only of order millivolts.

    Method: Link each fault to a remedy. Cleaning and conductive gel lower and stabilise skin-electrode contact resistance. Securing the electrodes and keeping the patient still reduces movement artefact. Shielded leads and separation from mains wiring reduce induced electrical noise. A differential, low-noise amplifier rejects common interference, while high gain makes the millivolt ECG large enough to record.

3.10.4.1 · Ultrasound imaging

  • Acoustic impedance is Z=ρcZ=\rho c. At a boundary, the reflected intensity fraction is Ir/Ii=((Z2Z1)/(Z2+Z1))2I_r/I_i=((Z_2-Z_1)/(Z_2+Z_1))^2.
  • Coupling gel removes the air layer between transducer and skin, reducing the impedance mismatch and allowing more ultrasound to enter the body.
  • A piezoelectric crystal driven by a short alternating potential difference produces an ultrasound pulse; a returning echo deforms the crystal and generates a potential difference.
  • Echo depth follows d=ct/2d=ct/2 because the measured time includes the outward and return journeys. Greater frequency improves resolution but generally increases attenuation.
  • An A-scan plots echo amplitude against time or depth, whereas a B-scan uses echo brightness and position to build a two-dimensional image.
  • Ultrasound is non-ionising and can image moving structures in real time, but it has lower resolution than some alternatives and is strongly reflected at air or bone boundaries.

Tier 1 · Easy

  1. 1. A display shows echo amplitude against time rather than a two-dimensional image. Identify the ultrasound scan type and state what a B-scan displays instead.[2 marks]

    Answer

    • It is an A-scan; a B-scan displays a two-dimensional brightness image.

    Method: An amplitude scan is one-dimensional, with echo size plotted against travel time or depth. A brightness scan places echoes spatially and represents their strength by brightness to form a two-dimensional image.

Tier 2 · Standard

  1. 1. Ultrasound passes normally from tissue A, where ρ=1000kg m3\rho=1000\,\text{kg m}^{-3} and c=1500m s1c=1500\,\text{m s}^{-1}, into tissue B, where ρ=1060kg m3\rho=1060\,\text{kg m}^{-3} and c=1540m s1c=1540\,\text{m s}^{-1}. Calculate the percentage of incident intensity reflected at the boundary.[4 marks]

    Answer

    • 0.179%0.179\%

    Method: The impedances are ZA=ρAcA=1000(1500)=1.500×106kg m2s1Z_A=\rho_Ac_A=1000(1500)=1.500\times10^6\,\text{kg m}^{-2}\text{s}^{-1} and ZB=1060(1540)=1.6324×106kg m2s1Z_B=1060(1540)=1.6324\times10^6\,\text{kg m}^{-2}\text{s}^{-1}. Thus Ir/Ii=((ZBZA)/(ZB+ZA))2=((1.63241.500)/(1.6324+1.500))2=1.787×103I_r/I_i=((Z_B-Z_A)/(Z_B+Z_A))^2=((1.6324-1.500)/(1.6324+1.500))^2=1.787\times10^{-3}. Multiplying by 100100 gives 0.179%0.179\%.

Tier 3 · Hard

  1. 1. A 4.0MHz4.0\,\text{MHz} ultrasound pulse travels through tissue at 1540m s11540\,\text{m s}^{-1} and returns to the transducer 130μs130\,\mu\text{s} after transmission. Calculate the reflector depth and estimate the scan resolution as one wavelength. Explain why coupling gel is placed between the transducer and skin.[6 marks]

    Answer

    • 0.100m0.100\,\text{m} depth and 0.39mm0.39\,\text{mm} resolution; the gel removes air and reduces the acoustic-impedance mismatch, so less intensity is reflected at the skin.

    Method: The pulse makes a round trip, so d=ct/2=1540(130×106)/2=0.1001md=ct/2=1540(130\times10^{-6})/2=0.1001\,\text{m}, giving 0.100m0.100\,\text{m}. Its wavelength is λ=c/f=1540/(4.0×106)=3.85×104m=0.385mm\lambda=c/f=1540/(4.0\times10^6)=3.85\times10^{-4}\,\text{m}=0.385\,\text{mm}, so a one-wavelength resolution estimate is 0.39mm0.39\,\text{mm}. Air has a very different acoustic impedance from tissue and would reflect most incident ultrasound; gel displaces the air and provides a better impedance match, increasing transmission into the body.

3.10.4.2 · Fibre optics and endoscopy

  • An optical fibre has a higher-index core surrounded by lower-index cladding, so rays incident above the critical angle undergo total internal reflection at the interface.
  • A coherent bundle keeps the fibres in the same relative positions at both ends and transfers an image; cladding prevents light crossing between fibres and blurring that image.
  • A non-coherent bundle does not preserve fibre order and carries illumination into the body, so its individual fibre positions do not encode an image.
  • Flexible fibre bundles permit internal imaging through a small opening and can guide illumination or treatment light with less invasive access.
  • For total internal reflection, light must travel from higher to lower refractive index and the incidence angle must exceed the critical angle; merely stating that light reflects is insufficient.

Tier 1 · Easy

  1. 1. Name the endoscope fibre bundle that carries an image and the bundle that carries illumination into the body.[2 marks]

    Answer

    • The coherent bundle carries the image, and the non-coherent (incoherent) bundle carries illumination.

    Method: Image information requires the fibre positions to be preserved, so it uses a coherent bundle. Illumination only needs light delivery and therefore uses a non-coherent or incoherent bundle.

Tier 2 · Standard

  1. 1. A fibre has core refractive index 1.621.62 and cladding refractive index 1.501.50. Calculate the critical angle at the core-cladding boundary and determine whether a ray incident there at 72.072.0^\circ undergoes total internal reflection.[3 marks]

    Answer

    • 67.867.8^\circ; total internal reflection occurs.

    Method: For light travelling from core to cladding, sinc=ncladding/ncore=1.50/1.62\sin c=n_{\text{cladding}}/n_{\text{core}}=1.50/1.62. Hence c=sin1(1.50/1.62)=67.8c=\sin^{-1}(1.50/1.62)=67.8^\circ. The incidence angle 72.072.0^\circ is greater than cc, and the light is travelling from higher to lower refractive index, so total internal reflection occurs.

Tier 3 · Hard

  1. 1. A damaged endoscope has an image bundle whose fibre positions are rearranged between its two ends, and some cladding has been removed. Explain the effects on the observed image and why the illumination bundle can still work when its fibre order is random.[5 marks]

    Answer

    • Rearranging the image fibres destroys the one-to-one spatial mapping, so light from each object point emerges at the wrong position. Missing cladding allows light to cross between neighbouring fibres or escape, reducing contrast and blurring the image. Random order does not prevent the illumination bundle working because it only delivers light into the body and does not encode spatial information.

    Method: A coherent image bundle must preserve the relative position of every fibre; losing that order scrambles the image. Core-cladding total internal reflection confines each ray, so missing cladding permits leakage and cross-talk, producing dimmer or blurred regions. In the illumination bundle, only the total delivered light matters, so a non-coherent arrangement remains suitable.

3.10.4.3 · Magnetic resonance (MR) scanner

  • A strong static magnetic field, produced by a superconducting magnet, makes hydrogen nuclei (protons) adopt aligned spin states and precess about the field direction; the field does not make stationary protons start spinning classically.
  • A short radio-frequency pulse at the appropriate frequency excites protons and changes their spin state in the selected region.
  • After the pulse, excited protons relax or de-excite and emit radio-frequency signals that receiver coils detect; detailed relaxation-time calculations are not required.
  • Gradient magnetic fields vary with position so successive small regions of a cross-section can be selected and located.
  • A computer processes the detected signals to construct a cross-sectional image. MR uses no ionising radiation and is particularly useful for soft-tissue contrast.

Tier 1 · Easy

  1. 1. State which nuclei provide the main signal in an MR scanner and describe their behaviour in the scanner's static magnetic field.[2 marks]

    Answer

    • Hydrogen nuclei (protons); their spins align and precess about the magnetic field direction.

    Method: The abundant hydrogen nuclei in body tissue supply the signal. In the strong static field, their spin states become aligned and the proton magnetic moments precess about the field lines.

Tier 2 · Standard

  1. 1. Describe what happens to selected protons during and immediately after a short radio-frequency pulse in an MR scan.[3 marks]

    Answer

    • The pulse excites the protons and changes their spin state. When the pulse stops, they relax or de-excite and emit radio-frequency signals that can be detected.

    Method: During the pulse, radio-frequency energy at the appropriate frequency is absorbed and selected protons change to an excited spin state. Once excitation ends, the protons return towards their original state. This relaxation or de-excitation releases an RF signal for the receiver coils.

Tier 3 · Hard

  1. 1. Outline how an MR scanner obtains a cross-sectional image, beginning with a patient in the main magnetic field and ending with a computer-generated image.[5 marks]

    Answer

    • The main field aligns proton spins, which precess about the field. Gradient coils select and locate successive small regions of the cross-section. Short RF pulses excite protons in each selected region; as they relax, the protons emit RF signals. Receiver coils detect the signals and a computer uses their position and strength to construct the image.

    Method: Give the process in causal order: alignment and precession in the static field; position-dependent gradient fields selecting a slice or small region; an RF pulse changing proton spin states; relaxation or de-excitation producing RF emissions; and detection followed by computer processing into a spatial image. No calculation of relaxation times is needed.

3.10.5.1 · The physics of diagnostic X-rays

  • Electrons released by thermionic emission are accelerated through a potential difference VV and decelerated at a metal target. Most input energy heats the target; a rotating anode spreads this heating over a larger area.
  • The continuous spectrum comes from electrons losing different amounts of kinetic energy. The maximum photon energy is Emax=eVE_{\max}=eV, so the minimum wavelength is λmin=hc/(eV)\lambda_{\min}=hc/(eV).
  • Characteristic lines arise when incident electrons remove inner-shell electrons and higher-shell electrons fall into the vacancies; their photon energies are differences between atomic energy levels.
  • Increasing tube current increases beam intensity and patient dose, whereas increasing accelerating voltage raises the maximum photon energy and penetration. A smaller focal spot improves sharpness but concentrates heating; filtration and collimation remove low-energy photons and restrict the exposed region.

Tier 1 · Easy

  1. 1. State the maximum photon energy, in keV\text{keV}, from an X-ray tube operating at 68kV68\,\text{kV}.[1 mark]

    Answer

    • 68keV68\,\text{keV}

    Method: An electron accelerated through 68kV68\,\text{kV} gains 68keV68\,\text{keV}. If it loses all this energy in one interaction, Emax=68keVE_{\max}=68\,\text{keV}.

Tier 2 · Standard

  1. 1. An X-ray tube uses an accelerating potential difference of 72kV72\,\text{kV}. Calculate the minimum X-ray wavelength. Use e=1.60×1019Ce=1.60\times10^{-19}\,\text{C}, h=6.63×1034J sh=6.63\times10^{-34}\,\text{J s} and c=3.00×108m s1c=3.00\times10^8\,\text{m s}^{-1}.[3 marks]

    Answer

    • 1.7×1011m1.7\times10^{-11}\,\text{m}

    Method: Convert the voltage: V=72×103VV=72\times10^3\,\text{V}. The maximum photon energy is Emax=eV=(1.60×1019)(72×103)=1.152×1014JE_{\max}=eV=(1.60\times10^{-19})(72\times10^3)=1.152\times10^{-14}\,\text{J}. Hence λmin=hc/Emax=(6.63×1034)(3.00×108)/(1.152×1014)=1.7266×1011m\lambda_{\min}=hc/E_{\max}=(6.63\times10^{-34})(3.00\times10^8)/(1.152\times10^{-14})=1.7266\times10^{-11}\,\text{m}, which is 1.7×1011m1.7\times10^{-11}\,\text{m} to two significant figures.

Tier 3 · Hard

  1. 1. A rotating-anode tube operates at 95kV95\,\text{kV} with a current of 3.0mA3.0\,\text{mA} for 0.080s0.080\,\text{s}. It converts 1.2%1.2\% of the electrical energy into X-rays. Calculate the maximum photon energy and the total X-ray energy produced. Explain one benefit of rotating the anode. Use e=1.60×1019Ce=1.60\times10^{-19}\,\text{C}.[5 marks]

    Answer

    • 1.5×1014J1.5\times10^{-14}\,\text{J} and 0.27J0.27\,\text{J}; rotation spreads heating over the target

    Method: The maximum photon energy is Emax=eV=(1.60×1019)(95×103)=1.52×1014J=1.5×1014JE_{\max}=eV=(1.60\times10^{-19})(95\times10^3)=1.52\times10^{-14}\,\text{J}=1.5\times10^{-14}\,\text{J} to two significant figures. The electrical energy supplied is VIt=(95×103)(3.0×103)(0.080)=22.8JVIt=(95\times10^3)(3.0\times10^{-3})(0.080)=22.8\,\text{J}. Therefore the X-ray energy is 0.012×22.8=0.2736J=0.27J0.012\times22.8=0.2736\,\text{J}=0.27\,\text{J}. Rotating the anode continually changes the impact area, spreading the large thermal energy and reducing local overheating or target damage.

3.10.5.2 · Image detection and enhancement

  • In a flat-panel detector, a scintillator converts X-rays to visible photons, photodiode pixels convert the light to electrical charge, and electronic scanning reads the pixel signals into a computer.
  • Flat-panel detectors are sensitive, produce an immediate digital image and allow computer enhancement, storage and transmission. Their wide dynamic range can reduce repeat exposures compared with photographic film.
  • X-rays darken photographic film. An intensifying screen converts X-rays to visible light so that less X-ray exposure is needed; fluoroscopic image intensification gives a live image but prolonged viewing can increase patient dose.
  • An X-ray-opaque contrast medium such as barium absorbs strongly. It outlines a low-contrast organ against surrounding soft tissue; a common error is to describe the image as being formed by reflected X-rays.

Tier 1 · Easy

  1. 1. State the function of the scintillator in a flat-panel X-ray detector.[1 mark]

    Answer

    • It converts incident X-ray photons into visible-light photons.

    Method: The marking point is the energy conversion: the scintillator absorbs X-rays and emits visible light, which the photodiode pixels can detect.

Tier 2 · Standard

  1. 1. Explain why a patient is given a barium suspension before an X-ray image of the digestive tract is recorded.[3 marks]

    Answer

    • Barium absorbs X-rays strongly, so less radiation reaches the detector behind the digestive tract and its outline has greater contrast with surrounding soft tissue.

    Method: Barium is X-ray opaque, so it has greater attenuation than the surrounding soft tissues. Regions behind the barium therefore give a different detector signal because fewer X-rays are transmitted. This increases image contrast and makes the shape of the digestive tract visible.

Tier 3 · Hard

  1. 1. A clinic needs a single abdominal X-ray image that can be enhanced immediately while keeping patient dose as low as practicable. Discuss the suitability of photographic film, fluoroscopic image intensification and a flat-panel detector, and justify a choice.[5 marks]

    Answer

    • Use the flat-panel detector: it is sensitive, gives an immediate digital image for enhancement and avoids the prolonged exposure of fluoroscopy; film needs processing and is less suitable for immediate enhancement.

    Method: Photographic film records a still image but needs chemical processing and is not directly available for computer enhancement. An intensifying screen can reduce the film exposure, but the result is still not immediate. Fluoroscopic image intensification is valuable for live motion, but continuous viewing exposes the patient throughout the procedure and is unnecessary for one still image. A flat-panel detector is sensitive, so a lower exposure can produce a usable signal, and electronic scanning supplies an immediate digital image that can be enhanced. Therefore the flat-panel detector best meets both stated requirements.

3.10.5.3 · Absorption of X-rays

  • For a narrow monoenergetic beam, transmitted intensity follows I=I0eμxI=I_0e^{-\mu x}, where μ\mu is the linear attenuation coefficient and xx is the material thickness. Use compatible units for μ\mu and xx.
  • The half-value thickness is x1/2=ln2/μx_{1/2}=\ln 2/\mu. After nn half-value thicknesses the intensity is I0(1/2)nI_0(1/2)^n; do not treat attenuation as a constant subtraction per unit thickness.
  • The mass attenuation coefficient is μm=μ/ρ\mu_m=\mu/\rho and has units m2kg1\text{m}^2\,\text{kg}^{-1} when SI units are used. It allows attenuation properties to be compared without the direct effect of density.
  • Bone and contrast media generally attenuate X-rays more strongly than soft tissue, producing differential detector signals. Examiners require transmitted, not absorbed, intensity when I=I0eμxI=I_0e^{-\mu x} is used.

Tier 1 · Easy

  1. 1. An absorber has linear attenuation coefficient 0.28cm10.28\,\text{cm}^{-1}. Calculate its half-value thickness.[2 marks]

    Answer

    • 2.5cm2.5\,\text{cm}

    Method: Use x1/2=ln2/μ=0.693/0.28=2.48cmx_{1/2}=\ln 2/\mu=0.693/0.28=2.48\,\text{cm}, which is 2.5cm2.5\,\text{cm} to two significant figures.

Tier 2 · Standard

  1. 1. A narrow X-ray beam of intensity 6.4W m26.4\,\text{W m}^{-2} passes through 7.0mm7.0\,\text{mm} of tissue with μ=0.18mm1\mu=0.18\,\text{mm}^{-1}. Calculate the transmitted intensity.[3 marks]

    Answer

    • 1.8W m21.8\,\text{W m}^{-2}

    Method: Use I=I0eμxI=I_0e^{-\mu x}. The exponent is μx=(0.18)(7.0)=1.26-\mu x=-(0.18)(7.0)=-1.26. Therefore I=6.4e1.26=1.82W m2=1.8W m2I=6.4e^{-1.26}=1.82\,\text{W m}^{-2}=1.8\,\text{W m}^{-2}.

Tier 3 · Hard

  1. 1. An incident X-ray beam has intensity 10.0W m210.0\,\text{W m}^{-2}. One ray crosses 18mm18\,\text{mm} of soft tissue with μ=0.050mm1\mu=0.050\,\text{mm}^{-1} and then 4.0mm4.0\,\text{mm} of bone with μ=0.32mm1\mu=0.32\,\text{mm}^{-1}. A neighbouring ray crosses 22mm22\,\text{mm} of the same soft tissue only. Calculate both transmitted intensities and the ratio of the soft-tissue-only intensity to the bone-path intensity.[5 marks]

    Answer

    • 1.1W m21.1\,\text{W m}^{-2}, 3.3W m23.3\,\text{W m}^{-2} and a ratio of 2.92.9

    Method: Attenuation exponents add for successive materials. For the bone path, μx=(0.050)(18)+(0.32)(4.0)=0.90+1.28=2.18\mu x=(0.050)(18)+(0.32)(4.0)=0.90+1.28=2.18, so Ibone=10.0e2.18=1.130W m2=1.1W m2I_{\text{bone}}=10.0e^{-2.18}=1.130\,\text{W m}^{-2}=1.1\,\text{W m}^{-2}. For soft tissue only, μx=(0.050)(22)=1.10\mu x=(0.050)(22)=1.10, so Isoft=10.0e1.10=3.329W m2=3.3W m2I_{\text{soft}}=10.0e^{-1.10}=3.329\,\text{W m}^{-2}=3.3\,\text{W m}^{-2}. Using unrounded values, the ratio is Isoft/Ibone=2.945=2.9I_{\text{soft}}/I_{\text{bone}}=2.945=2.9 to two significant figures.

3.10.5.4 · CT scanner

  • A CT scanner moves an X-ray tube around the patient and sends a narrow, approximately monochromatic beam through the body to an array of detectors at many angles.
  • Each detector measures transmitted intensity along a path. A computer combines many projections to reconstruct cross-sectional slices and can assemble successive slices into a three-dimensional representation.
  • CT removes the superposition of structures found in a plain radiograph and can give improved contrast and spatial information, especially for internal soft-tissue structures.
  • CT is more expensive and usually gives a larger ionising-radiation dose than a simple X-ray image. Comparisons must link resolution, cost and safety to the clinical task rather than merely list scanner features.

Tier 1 · Easy

  1. 1. State why the X-ray tube moves around a patient during a CT scan.[1 mark]

    Answer

    • To obtain attenuation measurements through the body from many directions.

    Method: Movement changes the beam direction, providing the many projections that the computer needs to reconstruct a cross-section.

Tier 2 · Standard

  1. 1. Describe how a CT scanner produces a cross-sectional image from X-rays transmitted through a patient.[3 marks]

    Answer

    • A narrow beam and detector array collect transmitted intensities at many angles as the tube moves, and a computer reconstructs these projection data into a cross-sectional image.

    Method: The moving tube directs a narrow X-ray beam through the selected section from a succession of angles. The detector array measures the transmitted intensity for many paths. A computer processes this set of projections to calculate the attenuation distribution and display a cross-sectional image.

Tier 3 · Hard

  1. 1. A hospital is choosing between CT and a plain X-ray image to localise a small lung lesion partly hidden by overlapping ribs, and to confirm the position of a radiopaque feeding tube. Compare the techniques using image resolution, superposition, radiation dose, cost and availability. Recommend one technique for each task.[6 marks]

    Answer

    • Use CT to localise the lung lesion because cross-sectional images remove rib superposition and give better depth information; use a plain X-ray for the feeding tube because its radiopaque position is shown adequately at lower dose, lower cost and greater availability.

    Method: CT reconstructs slices, so ribs at different depths do not remain superimposed on the lesion and its three-dimensional location is clearer. That extra information can justify CT's higher ionising dose, cost and lower availability for the lesion. A plain radiograph is quick, cheap, widely available and gives a lower dose. Because a radiopaque tube already has strong contrast and only its overall position must be checked, the projection image is sufficient. Therefore use CT for the obscured lesion and a plain X-ray for the tube check.

3.10.6.1 · Imaging techniques

  • A tracer contains a gamma-emitting radioisotope attached to a compound with an affinity for a particular organ. Gamma radiation is penetrating enough to leave the body and be detected externally.
  • Technetium-99m has a physical half-life of about 6h6\,\text{h} and emits a useful 140keV140\,\text{keV} gamma photon. Its short half-life limits dose while allowing a scan, and it can be labelled to many biologically active compounds.
  • Iodine-131 has a half-life of about 8d8\,\text{d}, emits beta radiation and a principal 364keV364\,\text{keV} gamma photon, and is naturally taken up by the thyroid. Indium-111 has a half-life of about 2.8d2.8\,\text{d}, emits 171keV171\,\text{keV} and 245keV245\,\text{keV} gamma photons, and can label compounds or blood cells for longer investigations.
  • A molybdenum-technetium generator contains longer-lived molybdenum-99, whose decay produces technetium-99m. It lets hospitals obtain fresh tracer without a nearby reactor or accelerator.
  • In PET, a positron-emitting tracer produces annihilation photons. Two 511keV511\,\text{keV} gamma photons travelling in nearly opposite directions are detected in coincidence, allowing the computer to locate tracer activity.

Tier 1 · Easy

  1. 1. State why a medical tracer used for external imaging should emit gamma radiation.[1 mark]

    Answer

    • Gamma radiation can pass out of the body and be detected externally.

    Method: The required property is penetration: enough gamma photons must escape the tissue to reach the detector outside the patient.

Tier 2 · Standard

  1. 1. Justify the use of technetium-99m for imaging an organ. Refer to its 6h6\,\text{h} half-life, its 140keV140\,\text{keV} gamma emission and its chemical use.[3 marks]

    Answer

    • Its half-life is long enough for preparation and imaging but short enough to limit dose; 140keV140\,\text{keV} gamma photons can leave the body for detection; and technetium-99m can be attached to compounds that concentrate in a chosen organ.

    Method: Award one linked justification for each stated property. A 6h6\,\text{h} half-life provides usable activity during the scan but causes activity to fall soon afterwards. The 140keV140\,\text{keV} gamma radiation is penetrating enough for external detection while avoiding an unnecessarily high photon energy. Labelling a compound with organ affinity makes the recorded count distribution show that organ's function or structure.

Tier 3 · Hard

  1. 1. Explain how a PET scan can map regions of high glucose uptake after a patient receives a positron-emitting glucose analogue. Include the nuclear event, the detection method and how position is inferred.[5 marks]

    Answer

    • The labelled glucose accumulates in active tissue; emitted positrons annihilate with electrons to produce two opposite 511keV511\,\text{keV} photons; coincidence detectors identify a line of response, and many such events are reconstructed into an uptake map.

    Method: The glucose analogue is carried into tissues, so regions with greater uptake contain more tracer. A radionuclide in the tracer emits a positron. After losing energy, the positron annihilates with an electron and produces two 511keV511\,\text{keV} gamma photons travelling in approximately opposite directions. Detectors on opposite sides register near-simultaneous photons as a coincidence, locating the event somewhere along the joining line. A computer combines many lines of response to reconstruct the spatial distribution of uptake.

3.10.6.2 · Half-life

  • Physical half-life TPT_P is the time for half the unstable nuclei to decay. It is a property of the radionuclide and is not altered by biological processes.
  • Biological half-life TBT_B is the time for biological removal processes to reduce the amount of tracer in an organ or body to half its value, ignoring radioactive decay.
  • Effective half-life TET_E describes the combined fall due to both processes and obeys 1/TE=1/TB+1/TP1/T_E=1/T_B+1/T_P. Consequently TET_E must be shorter than both TBT_B and TPT_P.
  • Keep all three half-lives in the same units before using the reciprocal equation. A common error is to add the half-lives themselves rather than their reciprocals.

Tier 1 · Easy

  1. 1. Define the biological half-life of a tracer in an organ.[2 marks]

    Answer

    • The time for biological processes alone to remove half of the tracer from the organ.

    Method: The definition must identify removal by biological processes and a reduction to one half; radioactive decay is excluded from this definition.

Tier 2 · Standard

  1. 1. A tracer has physical half-life 18h18\,\text{h} and biological half-life 30h30\,\text{h}. Calculate its effective half-life.[3 marks]

    Answer

    • 11h11\,\text{h}

    Method: Use 1/TE=1/TB+1/TP=1/30+1/18=0.0889h11/T_E=1/T_B+1/T_P=1/30+1/18=0.0889\,\text{h}^{-1}. Therefore TE=1/0.0889=11.25hT_E=1/0.0889=11.25\,\text{h}, which is 11h11\,\text{h} to two significant figures.

Tier 3 · Hard

  1. 1. A radionuclide has physical half-life 12.0h12.0\,\text{h}. Measurements in a patient give an effective half-life of 7.20h7.20\,\text{h}. Calculate the biological half-life and the percentage of the initial activity remaining in the patient after 21.6h21.6\,\text{h}.[5 marks]

    Answer

    • 18.0h18.0\,\text{h} and 12.5%12.5\%

    Method: Rearrange the reciprocal relation: 1/TB=1/TE1/TP=1/7.201/12.0=0.0556h11/T_B=1/T_E-1/T_P=1/7.20-1/12.0=0.0556\,\text{h}^{-1}, so TB=18.0hT_B=18.0\,\text{h}. The elapsed time is 21.6/7.20=3.0021.6/7.20=3.00 effective half-lives. The remaining activity fraction is therefore (1/2)3=0.125(1/2)^3=0.125, giving 12.5%12.5\%.

3.10.6.3 · Gamma camera

  • A lead collimator absorbs photons travelling in unsuitable directions, so accepted gamma photons carry directional information. Narrower holes improve spatial resolution but reduce count rate and sensitivity.
  • A scintillation crystal converts an absorbed gamma photon into a flash of visible light. A light guide spreads the flash to an array of photomultiplier tubes.
  • In a photomultiplier tube, light ejects photoelectrons from a photocathode. Successive dynodes at increasing potentials produce secondary emission, giving a large electron pulse at the anode.
  • The relative photomultiplier outputs locate each scintillation, while pulse size estimates photon energy so scattered photons can be rejected. A computer stores many accepted positions to build the image.

Tier 1 · Easy

  1. 1. State the function of the lead collimator in a gamma camera.[1 mark]

    Answer

    • It absorbs gamma photons arriving from unsuitable directions so that detected photons retain directional information.

    Method: The collimator selects photon directions before the scintillator; it does not focus gamma rays with refraction.

Tier 2 · Standard

  1. 1. Describe how a gamma photon entering a gamma camera produces an electrical pulse at the output of a photomultiplier tube.[3 marks]

    Answer

    • The photon produces a light flash in the scintillator, the photocathode emits photoelectrons, and successive dynodes multiply them to give an anode pulse.

    Method: The gamma photon deposits energy in the scintillation crystal, which emits visible photons. Light reaching a photomultiplier photocathode releases electrons by the photoelectric effect. These electrons accelerate to successive dynodes and cause secondary emission at each stage, so a much larger electron pulse is collected at the anode.

Tier 3 · Hard

  1. 1. Explain how a gamma camera determines the position of tracer activity and improves the quality of the recorded image. Include the roles of the collimator, photomultiplier array and pulse-height selection.[5 marks]

    Answer

    • The collimator selects photon directions; relative signals from nearby photomultipliers locate each scintillation; pulse height estimates photon energy so scattered events are rejected; accumulating accepted events produces the image, with a resolution-sensitivity trade-off set by the collimator.

    Method: The lead collimator accepts only a restricted range of directions, linking a detected photon to a line from the patient. A scintillation spreads light to several photomultipliers, and the relative sizes of their pulses allow the computer to estimate the interaction position. The total pulse height is related to deposited photon energy, so an energy window rejects lower-energy scattered photons that would be placed incorrectly. The computer accumulates the accepted positions to form the count image. Smaller collimator apertures sharpen directional selection and improve resolution, but fewer photons pass, so sensitivity falls.

3.10.6.4 · Use of high-energy X-rays

  • External radiotherapy directs high-energy X-rays into a tumour. Their penetration lets energy be delivered at depth, where ionisation damages DNA and prevents cells from dividing successfully.
  • Several beam directions can converge on the tumour. The tumour receives the full combined dose, while each region of healthy tissue lies in fewer beam paths and receives a smaller dose.
  • Collimation, shielding and computer planning restrict the irradiated volume. Imaging is used to locate the tumour before selecting beam directions and field shapes.
  • Treatment is divided into fractions so healthy cells can repair between exposures while the prescribed tumour dose accumulates. Do not claim that healthy tissue receives zero dose.

Tier 1 · Easy

  1. 1. State why high-energy X-rays are used to treat a tumour deep inside the body.[1 mark]

    Answer

    • They are sufficiently penetrating to reach and ionise the deep tumour.

    Method: The required link is high photon energy to penetration to the depth of the tumour.

Tier 2 · Standard

  1. 1. Explain how directing X-ray beams at a tumour from several angles can reduce damage to healthy tissue.[3 marks]

    Answer

    • All beams overlap at the tumour so their doses add there, whereas each healthy region is crossed by fewer beams and receives a smaller dose.

    Method: Aim each beam so that it passes through the tumour. At the tumour the energy deposited by all beam directions adds to the therapeutic dose. Away from the tumour, the entry and exit paths are spread through different healthy regions, so any one region receives only part of the total dose.

Tier 3 · Hard

  1. 1. A tumour of mass 0.18kg0.18\,\text{kg} receives 1.8Gy1.8\,\text{Gy} in each of 2020 X-ray treatment fractions. Calculate the total energy absorbed by the tumour. Explain two methods, other than reducing the prescribed tumour dose, that limit exposure of healthy cells.[5 marks]

    Answer

    • 6.5J6.5\,\text{J}; suitable methods include converging beams from different directions, collimation or shielding, and accurate imaging and planning

    Method: Since 1Gy=1J kg11\,\text{Gy}=1\,\text{J kg}^{-1}, the energy absorbed per fraction is E=Dm=(1.8)(0.18)=0.324JE=Dm=(1.8)(0.18)=0.324\,\text{J}. Over 2020 fractions, Etotal=20(0.324)=6.48J=6.5JE_{\text{total}}=20(0.324)=6.48\,\text{J}=6.5\,\text{J}. Healthy-tissue exposure can be limited by using several beam directions that overlap at the tumour but spread entry dose, and by collimating or shielding the beam to exclude other tissue. Accurate imaging and computer planning improve targeting so that less healthy tissue lies within the treatment fields.

3.10.6.5 · Use of radioactive implants

  • Brachytherapy places sealed radioactive sources inside or very close to a tumour, giving a high local dose while reducing irradiation of more distant tissue.
  • Beta-emitting implants are useful because beta particles ionise tissue over a short range. Gamma radiation would penetrate farther and expose more healthy tissue outside the target region.
  • Several small sources can distribute dose through an irregular tumour. Source activity, half-life, position and treatment time determine the delivered dose; temporary implants can be removed after the planned exposure.
  • Staff exposure is reduced using handling tools, shielding, short handling times and increased distance where appropriate. A common weak answer states only that beta is less penetrating without linking this to localisation of dose.

Tier 1 · Easy

  1. 1. State why a beta emitter is suitable for a radioactive implant placed inside a tumour.[1 mark]

    Answer

    • Beta radiation has a short range in tissue, so its ionising dose is concentrated near the implant.

    Method: Link the limited penetration of beta particles to a high local dose and reduced dose to distant healthy tissue.

Tier 2 · Standard

  1. 1. Explain why a beta-emitting implant may expose less healthy tissue than an external high-energy X-ray beam used to deliver the same tumour dose.[3 marks]

    Answer

    • The implant is beside the tumour and beta particles have a short range, so most energy is deposited locally; an external X-ray beam must cross healthy tissue before and after reaching the tumour.

    Method: The source is positioned within or adjacent to the target, avoiding a long entry path. Beta particles are strongly ionising over a limited range, so their energy is deposited close to the implant. High-energy X-rays are penetrating and an external beam necessarily irradiates some healthy tissue along its entry and exit paths.

Tier 3 · Hard

  1. 1. A temporary beta-emitting implant has initial activity 9.6MBq9.6\,\text{MBq} and physical half-life 48h48\,\text{h}. Calculate its activity after 120h120\,\text{h}. Explain how the position and radiation type of the implant help protect healthy tissue.[5 marks]

    Answer

    • 1.7MBq1.7\,\text{MBq}; placement in or beside the tumour and the short range of beta radiation localise the dose

    Method: The elapsed number of half-lives is 120/48=2.5120/48=2.5. Hence A=9.6(1/2)2.5=1.70MBq=1.7MBqA=9.6(1/2)^{2.5}=1.70\,\text{MBq}=1.7\,\text{MBq}. Placing the source within or close to the tumour avoids irradiating a long path of healthy tissue before the radiation reaches its target. Beta particles have a short range in tissue, so most ionisation occurs near the implant and relatively little dose reaches distant cells. The source can also be removed when the planned treatment time is complete.

3.10.6.6 · Imaging comparisons

  • A plain X-ray image is fast, widely available and gives good spatial detail for bone, but it uses ionising radiation, gives a two-dimensional projection and has poor soft-tissue contrast.
  • CT gives cross-sectional images with better localisation and soft-tissue information than a plain radiograph, but it usually has a higher ionising dose, higher cost and lower availability.
  • Ultrasound is portable, relatively inexpensive, real-time and non-ionising, but image quality is operator-dependent and transmission through bone or gas is poor. MR gives excellent soft-tissue contrast without ionising radiation but is slower, expensive and less available.
  • Radionuclide and PET images show physiological function rather than only anatomy, but require an internal ionising tracer and usually have lower spatial resolution. A comparison must apply resolution, dose or safety, convenience or availability, and tissue contrast to the stated clinical need.

Tier 1 · Easy

  1. 1. Select a suitable imaging technique for repeated bedside monitoring of moving soft tissue when ionising radiation must be avoided.[1 mark]

    Answer

    • Ultrasound imaging

    Method: Ultrasound is portable, produces real-time images and does not use ionising radiation, so it meets all three requirements.

Tier 2 · Standard

  1. 1. Compare CT and MR for repeated imaging of a brain tumour. Refer to soft-tissue contrast, patient dose and convenience.[3 marks]

    Answer

    • MR generally gives better soft-tissue contrast and no ionising dose, making it preferable for repeated scans, but CT is usually faster and more available while exposing the patient to ionising radiation.

    Method: MR distinguishes soft tissues well, so tumour boundaries can have high contrast. It uses magnetic fields and RF signals rather than ionising radiation, which is advantageous for repeated imaging. CT is often quicker and more readily available, but each scan contributes an X-ray dose; the justified preference is therefore MR unless speed or access is the overriding constraint.

Tier 3 · Hard

  1. 1. A patient may have a kidney stone. Evaluate ultrasound, CT and MR for locating the stone, using spatial resolution, soft-tissue or stone contrast, radiation dose, availability and convenience. Reach a justified recommendation.[6 marks]

    Answer

    • Ultrasound is a reasonable first choice because it is available, convenient and non-ionising; CT is preferred if a more precise high-resolution location is needed despite its larger dose; MR is less suitable because stone-like calcified material gives poor contrast and the scan is costly and less available.

    Method: Ultrasound is portable or widely accessible, quick and non-ionising, so it is suitable as an initial investigation, although its spatial resolution and image quality are lower and depend on the acoustic path. CT gives high-resolution cross-sectional localisation and strong contrast for dense stone material without superposition, but it uses a comparatively large ionising dose and is more costly. MR provides strong soft-tissue contrast without ionising radiation, yet calcified stone material is poorly shown, scans take longer and access is more limited. Therefore choose ultrasound first when it can answer the question, and use CT when the clinical need for precise localisation outweighs the radiation risk; MR is not the best choice for this target.