[2 marks]
Total for this question: 2
19 specification points · notes, questions, answers and worked methods
Checked against AQA 7408 section 3.10. Review basis: the qualification registry sourced from the AQA A-level Physics (7408) specification; registry verification recorded 11 July 2026.
Answer all questions in the spaces provided.
Explanation
Worked example
Explain why cone-rich vision gives greater spatial resolution than rod-rich vision.
Answer: Cones give greater spatial resolution because their low pathway convergence keeps signals from neighbouring retinal positions separate.
Common mistakes
Exam tip
A compare question should link receptor wiring to performance: greater convergence increases sensitivity but reduces spatial resolution.
[2 marks]
Total for this question: 2
[2 marks]
Total for this question: 2
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[4 marks]
Total for this question: 4
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[6 marks]
Total for this question: 6
[5 marks]
Total for this question: 5
Explanation
Worked example
A myopic eye has a far point from the eye. Calculate the power of a correcting lens that makes a distant object appear at the far point.
Answer: A diverging lens of power is required.
Common mistakes
Exam tip
State the sign and lens type with a calculated prescription; a negative dioptre value must be identified as diverging.
[2 marks]
Total for this question: 2
[2 marks]
Total for this question: 2
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[2 marks]
Total for this question: 2
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
Explanation
Worked example
Explain why the ossicles and oval window help transfer sound from air into inner-ear fluid.
Answer: The lever action and area reduction increase fluid pressure, improving transmission into the inner ear.
Common mistakes
Exam tip
For a transmission question, give the full ordered chain and name the change from mechanical motion to electrical signalling.
[2 marks]
Total for this question: 2
[2 marks]
Total for this question: 2
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[5 marks]
Total for this question: 5
[4 marks]
Total for this question: 4
[4 marks]
Total for this question: 4
[4 marks]
Total for this question: 4
[5 marks]
Total for this question: 5
Explanation
Worked example
Calculate the intensity level for .
Answer: The intensity level is .
Common mistakes
Exam tip
When comparing two sounds, use the intensity ratio directly in .
[2 marks]
Total for this question: 2
[2 marks]
Total for this question: 2
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[5 marks]
Total for this question: 5
[4 marks]
Total for this question: 4
[4 marks]
Total for this question: 4
[5 marks]
Total for this question: 5
[4 marks]
Total for this question: 4
Explanation
Worked example
At , a normal threshold is and an impaired threshold is . Determine the hearing loss.
Answer: The hearing loss at is .
Common mistakes
Exam tip
State both the affected frequency range and the upward change in required intensity level when interpreting hearing loss.
[1 mark]
Total for this question: 1
[1 mark]
Total for this question: 1
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[5 marks]
Total for this question: 5
[4 marks]
Total for this question: 4
[4 marks]
Total for this question: 4
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
Explanation
Worked example
Six R-to-R intervals occupy on an ECG trace. Calculate the heart rate.
Answer: The heart rate is .
Common mistakes
Exam tip
Measure several R-to-R intervals and divide by the number of intervals before converting period to beats per minute.
[2 marks]
Total for this question: 2
[2 marks]
Total for this question: 2
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[4 marks]
Total for this question: 4
[5 marks]
Total for this question: 5
[6 marks]
Total for this question: 6
Explanation
Worked example
An echo returns after emission. Calculate the boundary depth for .
Answer: The boundary is approximately deep.
Common mistakes
Exam tip
For an echo calculation, state the factor of two explicitly; for a comparison, balance non-ionising real-time imaging against attenuation and resolution.
[2 marks]
Total for this question: 2
[2 marks]
Total for this question: 2
[4 marks]
Total for this question: 4
[3 marks]
Total for this question: 3
[4 marks]
Total for this question: 4
[6 marks]
Total for this question: 6
[6 marks]
Total for this question: 6
[6 marks]
Total for this question: 6
[5 marks]
Total for this question: 5
[6 marks]
Total for this question: 6
Explanation
Worked example
Explain why an endoscope needs a coherent fibre bundle for imaging but may use a non-coherent bundle for illumination.
Answer: The coherent bundle transfers the image pattern; the non-coherent bundle can supply illumination.
Common mistakes
Exam tip
A full total-internal-reflection statement must name travel from higher to lower refractive index and incidence above the critical angle.
[2 marks]
Total for this question: 2
[2 marks]
Total for this question: 2
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[4 marks]
Total for this question: 4
[4 marks]
Total for this question: 4
[5 marks]
Total for this question: 5
Explanation
Worked example
Outline how an MR scanner obtains a cross-sectional image after the patient is placed in the main magnetic field.
Answer: Spatial selection by gradients, RF excitation, RF detection and computer processing produce the cross-sectional image.
Common mistakes
Exam tip
For an outline question, keep the sequence explicit: align and precess, select, excite, detect, then process.
[2 marks]
Total for this question: 2
[2 marks]
Total for this question: 2
[3 marks]
Total for this question: 3
[2 marks]
Total for this question: 2
[2 marks]
Total for this question: 2
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[4 marks]
Total for this question: 4
[6 marks]
Total for this question: 6
[5 marks]
Total for this question: 5
Explanation
Worked example
Calculate the maximum photon energy from a tube operating at , in joules and electronvolts.
Answer: .
Common mistakes
Exam tip
In a controls question, link current to intensity, voltage to maximum energy, focal spot to sharpness and filtration or collimation to dose.
[1 mark]
Total for this question: 1
[1 mark]
Total for this question: 1
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[5 marks]
Total for this question: 5
[3 marks]
Total for this question: 3
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[6 marks]
Total for this question: 6
Explanation
Worked example
Explain how a flat-panel detector converts an incident X-ray pattern into a digital image.
Answer: The scintillator-photodiode-scanning chain converts spatial X-ray intensity into digital pixel values.
Common mistakes
Exam tip
For an operation question, give the detector stages in order and identify the energy conversion at each stage.
[1 mark]
Total for this question: 1
[1 mark]
Total for this question: 1
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[6 marks]
Total for this question: 6
[4 marks]
Total for this question: 4
Explanation
Worked example
An absorber has . Calculate its half-value thickness.
Answer: The half-value thickness is .
Common mistakes
Exam tip
Check whether the question requests transmitted intensity, absorbed intensity or half-value thickness before selecting the relation.
[2 marks]
Total for this question: 2
[2 marks]
Total for this question: 2
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[5 marks]
Total for this question: 5
[4 marks]
Total for this question: 4
[4 marks]
Total for this question: 4
[5 marks]
Total for this question: 5
[4 marks]
Total for this question: 4
Explanation
Worked example
Explain why CT can distinguish overlapping internal structures better than a plain radiograph.
Answer: CT removes projection overlap by reconstructing a spatially resolved cross-section from many angular measurements.
Common mistakes
Exam tip
A comparison must connect the reconstructed slice to improved localisation, then balance that benefit against cost and ionising dose.
[1 mark]
Total for this question: 1
[1 mark]
Total for this question: 1
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[6 marks]
Total for this question: 6
[4 marks]
Total for this question: 4
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[4 marks]
Total for this question: 4
Explanation
Worked example
Explain why technetium-99m is suitable for diagnostic tracer imaging.
Answer: Detectable gamma emission, a suitably short half-life, flexible labelling and generator availability make technetium-99m useful.
Common mistakes
Exam tip
Justify a tracer by linking radiation type, half-life, gamma energy and organ-specific labelling to the intended scan.
[1 mark]
Total for this question: 1
[1 mark]
Total for this question: 1
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[6 marks]
Total for this question: 6
Explanation
Worked example
A tracer has and . Calculate .
Answer: The effective half-life is .
Common mistakes
Exam tip
Use the shorter-than-both check immediately after evaluating an effective half-life.
[2 marks]
Total for this question: 2
[2 marks]
Total for this question: 2
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[4 marks]
Total for this question: 4
[5 marks]
Total for this question: 5
[6 marks]
Total for this question: 6
Explanation
Worked example
Explain the trade-off produced by making gamma-camera collimator holes narrower.
Answer: Narrower holes improve spatial resolution but reduce sensitivity and require longer counting.
Common mistakes
Exam tip
An operation answer should follow one photon through collimator, crystal, photocathode, dynodes and anode before explaining image formation.
[1 mark]
Total for this question: 1
[1 mark]
Total for this question: 1
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[5 marks]
Total for this question: 5
[4 marks]
Total for this question: 4
[5 marks]
Total for this question: 5
[4 marks]
Total for this question: 4
[6 marks]
Total for this question: 6
Explanation
Worked example
Explain how using three beam directions reduces harm compared with delivering the entire tumour dose through one path.
Answer: Beam convergence maintains tumour dose while spreading the unavoidable healthy-tissue dose across different regions.
Common mistakes
Exam tip
Use comparative dose language: full combined dose at the tumour, smaller dose in each healthy-tissue path.
[1 mark]
Total for this question: 1
[1 mark]
Total for this question: 1
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[4 marks]
Total for this question: 4
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
Explanation
Worked example
Explain why a beta-emitting implant may be preferred to a gamma-emitting implant for a small local tumour.
Answer: The beta implant concentrates an effective ionising dose within the local tumour while limiting distant exposure.
Common mistakes
Exam tip
For a justify question, connect radiation range directly to the required target size and protection of surrounding tissue.
[1 mark]
Total for this question: 1
[1 mark]
Total for this question: 1
[3 marks]
Total for this question: 3
[2 marks]
Total for this question: 2
[3 marks]
Total for this question: 3
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[4 marks]
Total for this question: 4
[6 marks]
Total for this question: 6
[6 marks]
Total for this question: 6
Explanation
Worked example
Compare ultrasound and CT for repeated imaging of a moving fetus.
Answer: Ultrasound is preferred because it gives convenient real-time imaging without ionising dose.
Common mistakes
Exam tip
Structure comparisons as paired consequences for the stated case: resolution or contrast, convenience, then ionising-radiation risk.
[1 mark]
Total for this question: 1
[1 mark]
Total for this question: 1
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[3 marks]
Total for this question: 3
[6 marks]
Total for this question: 6
[5 marks]
Total for this question: 5
[5 marks]
Total for this question: 5
[6 marks]
Total for this question: 6
[6 marks]
Total for this question: 6
Answers begin on a new printed page so the question pack can be completed without the solutions alongside it.
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Low light levels stimulate the more sensitive rod cells. Rods do not provide colour vision, and the convergence of several rods onto one nerve pathway gives low spatial resolution; either image property earns the second marking point. | 2 |
| 02.1 |
| Cones operate best in bright light and provide colour vision. Their pathways show little convergence, so signals from neighbouring retinal positions remain separate. | 2 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Identify convergence: several rods connect to a common sensory pathway. Their signals are combined, which helps a small light stimulus reach the detection threshold. Because the same pathway does not identify which individual rod responded, the two source positions cannot be resolved. | 3 |
| 02.1 |
| Moving the object closer makes the rays reaching the eye more divergent. Award an equivalent accommodation description using ciliary-muscle contraction, increased lens curvature, increased power or decreased focal length when it is linked correctly. The additional refraction brings the rays to a focus at the fixed retinal surface. | 3 |
| 03.1 |
| Correct the optical sequence: the air-cornea boundary provides most of the refraction, and the lens provides further adjustable refraction. The emerging rays must actually converge at the retinal surface, so the image there is real. A converging eye system forms an inverted retinal image, not an upright virtual one. | 4 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| For bright light, award the linked chain cones, colour vision, and high resolution from separate or weakly convergent pathways. For dim light, award rods, no colour, and reduced resolution from convergence. Complete the comparison by linking that convergence to greater sensitivity because signals from several rods can combine. | 5 |
| 02.1 |
| For the faint object, link peripheral rod vision to convergence and summation, which increase sensitivity. For the close pair, link direct cone-rich vision to low convergence, so signals from neighbouring image positions are not combined. This preserves positional information and allows the pair to be resolved. | 5 |
| 03.1 |
| Separate the information encoded by receptor type from that encoded by pathway position. Preserved cone types retain the relative responses needed for the colour of a large uniform area. Randomised cone positions destroy the spatial mapping needed to reproduce fine detail. In very low light rods dominate; because their sensitivity and convergent pathways are unchanged, detection of the faint grey patch is not impaired by this cone-pathway fault. | 5 |
| 04.1 |
| For X, compare the single-cone input directly with threshold: , so neither separate pathway fires. For Y, nine rods of the group are illuminated, giving . Rod convergence therefore increases sensitivity. Because both images feed one pathway, the output contains no separate positional signals, so spatial resolution is lost. | 6 |
| 05.1 |
| Threshold sensitivity is inversely proportional to the intensity needed for detection, so . Therefore . Near threshold, the response is rod-dominated: rods do not encode colour and their convergent pathways combine positional information. Increased illumination activates cones; the different cone responses support colour vision, while little pathway convergence preserves fine spatial detail. | 5 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| A myopic eye is too powerful, so parallel rays would focus before the retina. A diverging lens reduces the convergence before the rays enter the eye; its focal length and power are negative. | 2 |
| 02.1 |
| Astigmatism involves unequal focusing in different planes. A cylindrical component supplies direction-dependent power, and its axis from to specifies the orientation of the correction. | 2 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| The focal length is . Hence , so . The magnification magnitude is . | 3 |
| 02.1 |
| Use , so to significant figures. The negative power identifies a diverging lens. It corrects myopia by making light from a distant object appear to come from a virtual image at the eye's far point. | 3 |
| 03.1 |
| Interpret the prescription rather than combining its numerical values. The sphere provides the same correction in all planes. The cylinder adds direction-dependent correction in the plane perpendicular to the stated axis, while the axis specifies the orientation with no cylindrical power. | 2 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 | The object is real, so . The required image is virtual, so . Then . Therefore , which is to significant figures. The positive sign confirms that a converging lens is required. | 5 | |
| 02.1 |
| For the distant sign, the lens must form a virtual image at the far point: and . Hence to significant figures. For the phone, , so . This virtual image lies between and in front of the eye, so the same lens allows the phone to be focused. | 5 |
| 03.1 |
| For M, a distant object has and must form a virtual image at the far point, so . Hence , giving and therefore a diverging lens. For H, and the required virtual image is at the near point, so . Thus , giving and therefore a converging lens. | 5 |
| 04.1 |
| At the closest clear position the lens must form a virtual image at the unaided near point, so . Using , and . The magnification magnitude is . The positive-power lens adds convergence, allowing a nearer object to produce a virtual image at a distance the eye can focus. | 5 |
| 05.1 |
| For a distant object, and . The contact lens is at the cornea, so and . The far point is in front of the spectacle lens, so and . Both lenses are diverging; the smaller lens-to-far-point distance for the spectacles ( against ) requires the slightly larger magnitude of power. | 5 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Sound pressure variations act on the tympanic membrane. Its vibration is transmitted through the malleus, incus and stapes, collectively called the ossicles. | 2 |
| 02.1 | Use . Therefore to significant figures. | 2 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| First identify the lever action of the ossicles and link it to an increased force. Then compare areas: the oval window is smaller than the tympanic membrane. Using , a larger force acting over a smaller area gives a greater pressure in the inner-ear fluid. | 3 |
| 02.1 |
| The ossicles transmit a forced vibration, so the oval window vibrates at the driving frequency. Its displacement amplitude is reduced, but the ossicle lever action and the smaller oval-window area increase the pressure delivered to the inner-ear fluid. | 3 |
| 03.1 |
| Follow the three credited stages in order: vibration of the tympanic membrane is passed through the ossicles to the oval window; the oval window produces a pressure wave in the inner-ear fluid; receptors are stimulated and produce electrical impulses in the auditory nerve. | 3 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 | The force on the tympanic membrane is . The ossicle output force is . Therefore to significant figures. | 5 | |
| 02.1 |
| The tympanic force is . The required oval-window force is . Hence the required force gain is to significant figures. The area ratio is , so the area reduction contributes much more than the lever gain. | 4 |
| 03.1 |
| For air conduction, trace the missing middle-ear link: the tympanic membrane still vibrates, but without the ossicles it does not drive the oval window effectively. The reduced inner-ear pressure wave gives less receptor stimulation and fewer auditory-nerve impulses. For bone conduction, vibration passes through the skull to the inner ear, bypassing the tympanic membrane-to-oval-window route and giving the distinct fourth credit point. | 4 |
| 04.1 |
| Use the relationship qualitatively: for the same force amplitude, spreading the force over a larger area reduces pressure. The oversized piston therefore produces a lower-amplitude pressure wave in the inner-ear fluid. This gives less receptor stimulation, fewer auditory-nerve impulses and a quieter perceived sound. | 4 |
| 05.1 |
| The force delivered to the fluid is . Hence to significant figures. A forced vibration retains the driving frequency, so the fluid wave is at . This mechanical pressure wave stimulates sensory receptors, producing electrical impulses in the auditory nerve. | 5 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 | Use . Thus . | 2 | |
| 02.1 | For an isotropic source, . Therefore , which is to significant figures. | 2 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Rearrange to give . Hence to significant figures. Using the unrounded intensity, to significant figures. | 3 |
| 02.1 |
| Intensity follows the inverse-square law, so . Hence . The level therefore decreases by to significant figures; is also acceptable. | 3 |
| 03.1 |
| Apply the supplied weighting to each unweighted level: and . The difference is not a physical-intensity difference; it results from frequency weighting that approximates the greater human sensitivity at . | 3 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| First find the intensity: . For an isotropic source, , so , giving to significant figures. Equal-loudness curves have their minimum near a few kilohertz; at a larger level is required for the same perceived loudness. | 5 |
| 02.1 |
| The level difference is . Therefore to significant figures. Since , equal receiving areas give the same power ratio. Requiring times the intensity for equal loudness shows that hearing is less sensitive at than at . | 4 |
| 03.1 |
| For the low tone, . For the high tone, . A level of at has the same loudness as about at . This exceeds the actual protected level by about , so the tone is perceived as louder. | 4 |
| 04.1 |
| Add intensities, not decibel readings. Relative to the same reference intensity, . Thus . Apply the supplied weighting: , which is below . | 5 |
| 05.1 |
| Equal loudness requires , so . Now and . Therefore to significant figures. The smaller high-frequency intensity is sufficient because the ear is more sensitive at than at . | 4 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Age-related hearing loss does not raise the threshold equally at all frequencies. The characteristic change is a larger loss towards the high-frequency end of an equal-loudness or threshold graph. | 1 |
| 02.1 |
| A localised loss or notch near is characteristic of deterioration associated with prolonged excessive-noise exposure. | 1 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| The hearing loss is the vertical threshold difference: . Using gives . The impaired threshold intensity is therefore times the normal value. | 3 |
| 02.1 |
| Hearing loss is the level difference, so . This is to significant figures. The impaired ear therefore needs a level higher for the same loudness, so its equal-loudness curve lies above the normal curve at that frequency. | 3 |
| 03.1 |
| Age-related deterioration is frequency dependent rather than a single threshold shift. The high-frequency part of the equal-loudness curve moves upward more, meaning that a larger added level is required there. Adding the same level everywhere leaves the difference between low- and high-frequency hearing losses uncompensated. | 3 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Compare the frequency dependence rather than the total loss. A's largest value is at , matching age-related high-frequency loss. B has a pronounced maximum at , matching excessive-noise exposure. A positive hearing loss raises the required level on the equal-loudness curve; the larger the loss, the larger the upward displacement and the quieter an unchanged sound is perceived. | 5 |
| 02.1 |
| The threshold ratio is . Therefore the hearing loss is . This produces a local upward displacement of about in the threshold curve at , while the neighbouring parts remain close to normal. A pronounced local notch around indicates excessive-noise deterioration rather than the broad high-frequency loss typical of ageing. | 4 |
| 03.1 |
| Use . For the normal curve, , so the ratio is . For the damaged-hearing curve, the difference is , giving . Equal loudness therefore changes from needing less intensity at to needing much more, showing frequency-dependent hearing loss around . | 4 |
| 04.1 |
| Subtract the aid gain from the impaired level required at each frequency. The external requirements are and . Compare with the normal curve: the residual losses are and . Because the residual shift is frequency dependent and larger at , the aid has not restored the original equal-loudness relationship. | 5 |
| 05.1 |
| The combined intensity factor is . Therefore . The level increase is , giving . The broad age-related loss raises the high-frequency region, while multiplying by the extra local factor produces a superimposed notch around . | 5 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Match each named feature to the electrical activity of the ventricles: rapid ventricular depolarisation produces QRS, and the later recovery or repolarisation produces T. | 2 |
| 02.1 |
| The P wave is produced by depolarisation of the atria. Skin electrodes measure a potential difference arising from the heart's electrical activity, so the trace is not a record of blood pressure. | 2 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Use for the horizontal paper scale. The PR interval is . Since , the measured conduction interval is within the supplied normal range. | 3 |
| 02.1 | The cardiac period is . The paper distance for one period is . This gives an R-peak separation of to significant figures. | 3 | |
| 03.1 |
| The P wave represents atrial depolarisation, so it is unaffected. The QRS complex represents ventricular depolarisation, so its timing is also unchanged. The T wave represents ventricular repolarisation and is therefore delayed. The R-R separation measures the cardiac period; the stated unchanged period keeps that separation constant. | 3 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Link each fault to a remedy. Cleaning and conductive gel lower and stabilise skin-electrode contact resistance. Securing the electrodes and keeping the patient still reduces movement artefact. Shielded leads and separation from mains wiring reduce induced electrical noise. A differential, low-noise amplifier rejects common interference, while high gain makes the millivolt ECG large enough to record. | 5 |
| 02.1 |
| Convert the input to volts: . The gain is . For one interval, the percentage uncertainty is . For five intervals it is . The same absolute timing uncertainty is a smaller fraction of the longer total time, so averaging over several intervals gives a more precise period and heart rate. | 5 |
| 03.1 |
| The differential heart signal is , so the output is . The common appears in both inputs and subtracts to zero. An additional at B changes by , so the output error is . Low, stable contact resistance from conductive gel reduces differences between electrode contacts and movement-related changes. | 4 |
| 04.1 |
| Convert millivolts to volts and multiply by gain: , and . Only QRS exceeds , so its peak amplitude is clipped and cannot be measured correctly. The largest input sets the gain limit: , so a gain no greater than about is required. | 5 |
| 05.1 |
| Six R peaks contain five R-R intervals. Their total time is , so the mean period is and the rate is . The voltage calibration gives . On a normal trace the broad feature after QRS is T, produced by ventricular repolarisation. | 6 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Ultrasound echoes can be processed continuously to display fetal motion in real time. Ultrasound is non-ionising, so it avoids the cumulative ionising-radiation risk that would make repeated X-ray imaging unsuitable. | 2 |
| 02.1 |
| An alternating potential difference makes the crystal deform and emit an ultrasound pulse. A returning echo deforms the crystal and produces a detected potential difference. | 2 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 | The impedances are and . Thus . Multiplying by gives . | 4 | |
| 02.1 |
| The extra round-trip time through the layer is . Hence the thickness is . This is to significant figures; accept . | 3 |
| 03.1 |
| The transducer supplies one line of pulse-echo data at each position. The computer converts each return time using , so the echo is placed at the correct depth. Its amplitude controls the displayed brightness. Placing successive lines beside one another at their transducer positions builds the two-dimensional B-scan. | 4 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| The pulse makes a round trip, so , giving . Its wavelength is , so a one-wavelength resolution estimate is . Air has a very different acoustic impedance from tissue and would reflect most incident ultrasound; gel displaces the air and provides a better impedance match, increasing transmission into the body. | 6 |
| 02.1 |
| Use . For boundary 1, . For boundary 2, . Since the incident intensities are equal, , giving to significant figures. Boundary 2 therefore gives the stronger echo. Its larger reflection coefficient means less ultrasound is transmitted through the bone, so echoes from structures beyond it are weak. | 6 |
| 03.1 |
| At , , so the resolution requirement is not met. Its round-trip attenuation is . At , , and the round-trip attenuation is . This is below while the wavelength is below , so only satisfies both constraints. | 6 |
| 04.1 |
| For the near face, . For the far face, . The pulse crosses the near face twice, so its returning fraction is . The ratio , so the far-boundary echo has the greater intensity and is displayed more brightly. | 5 |
| 05.1 |
| The latest echo makes a round trip of , so . The interval between pulses must be at least this long. Therefore . At a higher repetition frequency, a new pulse is transmitted before the deepest echo returns, so the receiver could associate that late echo with the newer pulse and calculate an incorrect depth. | 6 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Image information requires the fibre positions to be preserved, so it uses a coherent bundle. Illumination only needs light delivery and therefore uses a non-coherent or incoherent bundle. | 2 |
| 02.1 |
| Total internal reflection is possible only for travel from higher to lower refractive index. It occurs when the incidence angle at that boundary is greater than the critical angle. | 2 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| For light travelling from core to cladding, . Hence . The incidence angle is greater than , and the light is travelling from higher to lower refractive index, so total internal reflection occurs. | 3 |
| 02.1 |
| Credit the separate functions: an ordered coherent bundle carries the image back to the clinician, while fibres can carry light into the body for illumination or treatment. Because both bundles are narrow, they can reach the target through a small opening. This gives less invasive access, reducing damage to surrounding tissue. | 3 |
| 03.1 |
| For a meridional ray, increasing the angle to the fibre axis decreases the incidence angle at the side wall. Confinement requires incidence from the higher-index core at an angle greater than the critical angle. Here that condition fails, so the ray is transmitted into the lower-index cladding instead of being totally internally reflected. | 3 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| A coherent image bundle must preserve the relative position of every fibre; losing that order scrambles the image. Core-cladding total internal reflection confines each ray, so missing cladding permits leakage and cross-talk, producing dimmer or blurred regions. In the illumination bundle, only the total delivered light matters, so a non-coherent arrangement remains suitable. | 5 |
| 02.1 |
| For X, , so and total internal reflection does not occur. For Y, , so and the ray is confined. Total internal reflection transports light within each fibre, but the complete image is reproduced only if the fibres preserve their relative ordering from one end of the bundle to the other. | 5 |
| 03.1 |
| The image scale is , so the lesion image is wide. Bundle A needs , so it fails the sampling criterion. Bundle B needs only and therefore samples finely enough. At least one mark requires the separate coherence verdict: because B is non-coherent, it cannot preserve relative positions and cannot form an image. | 4 |
| 04.1 |
| At the core-cladding boundary, . The side-wall incidence angle and the ray angle to the axis sum to , so the limiting internal angle is . At the flat end, Snell's law gives , so . Entrance angles below this limit lead to incidence above the critical angle and total internal reflection. | 4 |
| 05.1 |
| Successive retained fractions multiply. With intact cladding, . With damage, . Their ratio is . Good confinement keeps each signal bright and within its own fibre, but it does not determine where that fibre ends; a coherent bundle must also keep relative fibre positions unchanged so the emerging pattern reproduces the image. | 5 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| The abundant hydrogen nuclei in body tissue supply the signal. In the strong static field, their spin states become aligned and the proton magnetic moments precess about the field lines. | 2 |
| 02.1 |
| A short radio-frequency pulse supplies energy to change proton spin states. Radio waves are non-ionising, unlike X-rays or gamma radiation. | 2 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| During the pulse, radio-frequency energy at the appropriate frequency is absorbed and selected protons change to an excited spin state. Once excitation ends, the protons return towards their original state. This relaxation or de-excitation releases an RF signal for the receiver coils. | 3 |
| 02.1 |
| De-excitation releases photons in the radio-frequency part of the electromagnetic spectrum. Their frequency is the same as that of the incident radio-frequency pulse. | 2 |
| 03.1 |
| Award one mark for correcting the classical-spin misconception: proton spin already exists and the main field aligns the spin axes. Award one mark for the spatial-selection step: gradient fields select successive small regions rather than the whole cross-section being scanned at once. | 2 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Give the process in causal order: alignment and precession in the static field; position-dependent gradient fields selecting a slice or small region; an RF pulse changing proton spin states; relaxation or de-excitation producing RF emissions; and detection followed by computer processing into a spatial image. No calculation of relaxation times is needed. | 5 |
| 02.1 |
| Separate signal production from image location. The static field, RF excitation, de-excitation and RF detection still operate, so a combined signal may be recorded. Gradient coils normally impose a position-dependent field to select and encode regions. With no gradient, the computer cannot map signal strength to coordinates in the patient. | 5 |
| 03.1 |
| Credit the MR-internal chain: the short pulse supplies energy and changes selected proton spin states; relaxation produces the RF response detected for the image; the nuclei involved are hydrogen nuclei (protons), abundant in body tissue; and the fields and radio waves used are non-ionising, so several scans do not accumulate a risk from tissue ionisation. | 4 |
| 04.1 |
| For B, . The coil emf for A is , so its output is . The difference is greater than the required , so the two shades are distinguishable. Gradient fields vary with position, selecting or encoding successive small regions; the computer combines that position information with the detected strengths to construct the cross-sectional image. | 6 |
| 05.1 |
| Correct each causal step. The static main field is established before RF excitation and aligns the proton spin axes. The RF transmitter, not the receiver coils, transfers energy to selected protons; the receiver coils detect the response during relaxation. De-excitation produces RF emission at the same frequency as the exciting pulse. Gradient fields encode position, and the computer combines that spatial information with signal amplitude to reconstruct the cross-sectional image. | 5 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 | An electron accelerated through gains . If it loses all this energy in one interaction, . | 1 | |
| 02.1 |
| A larger current sends more electrons to the target each second, so more X-ray photons are produced per second and the beam intensity increases. | 1 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 | Convert the voltage: . The maximum photon energy is . Hence , which is to two significant figures. | 3 | |
| 02.1 |
| Low-energy photons are more likely to be absorbed near the patient's surface and are unlikely to reach the detector. Aluminium filtration removes many of these photons before they enter the patient. The remaining beam has a greater mean photon energy, and the avoidable skin dose is reduced. | 3 |
| 03.1 | The photon energy equals the level separation. First, . Since , the separation is . | 3 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| The maximum photon energy is to two significant figures. The electrical energy supplied is . Therefore the X-ray energy is . Rotating the anode continually changes the impact area, spreading the large thermal energy and reducing local overheating or target damage. | 5 |
| 02.1 |
| The maximum X-ray photon energy is , so it depends on the tube potential difference rather than the focal-spot size. A fixed current means the same charge, and therefore the same number of electrons, reaches the anode each second. Reducing the impact area does not change that electron rate, so the photon count per second is approximately unchanged. | 3 |
| 03.1 |
| Since , . The X-ray power is , while . Substituting gives , so ; explicitly, the rates are and . Characteristic energies are fixed by tungsten's electron-level separations, so changing the voltage does not move those lines once both voltages exceed the relevant thresholds. | 5 |
| 04.1 |
| The photon energy is the difference between the binding energies: . In joules this is . Hence , giving to three significant figures. An incident electron must supply at least the K-shell binding energy, requiring a minimum tube potential difference of . Therefore is insufficient even though it exceeds the energy of the emitted photon. | 5 |
| 05.1 |
| The focal-spot area is . Its maximum heating power is . Since the heating fraction is , , so to two significant figures. The maximum photon energy is . Doubling the spot area doubles the allowed heating power and therefore the allowed current. A larger current supplies more electrons per second and increases photon intensity, but the unchanged potential difference leaves the energy gained by each electron, and hence , unchanged. | 6 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| The marking point is the energy conversion: the scintillator absorbs X-rays and emits visible light, which the photodiode pixels can detect. | 1 |
| 02.1 |
| Visible photons from the scintillator produce charge in the photodiode, giving an electrical signal for that pixel. | 1 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Barium is X-ray opaque, so it has greater attenuation than the surrounding soft tissues. Regions behind the barium therefore give a different detector signal because fewer X-rays are transmitted. This increases image contrast and makes the shape of the digestive tract visible. | 3 |
| 02.1 |
| The screen absorbs an X-ray photon and emits many visible photons. Photographic film responds much more strongly to this visible light than it does directly to X-rays, so the required film darkening is obtained with fewer incident X-ray photons and a smaller patient dose. The visible light spreads before reaching the film, making boundaries less sharp. | 3 |
| 03.1 |
| A scintillator and photodiode array convert the local X-ray intensity into a separate electrical charge at each position. Summing first retains only total exposure and discards the mapping between signal and position. The array must instead be addressed pixel by pixel during electronic scanning, so the computer can place each value at the corresponding image location and enhance the resulting digital pattern. | 3 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| The failed width is . Since the lesion lies wholly inside this wider strip, every photodiode position that would record it has failed, so the lesion produces no recorded spatial data. Digital enhancement can remap or interpolate recorded neighbouring values, but it cannot determine the unrecorded anatomy inside the strip and therefore cannot reconstruct the lesion. Repeating the exposure adds ionising-radiation dose, so it is justified only when the missing anatomy could affect diagnosis and the expected diagnostic benefit outweighs that added risk. | 5 |
| 02.1 |
| Barium is much more X-ray opaque than the surrounding soft tissues. Less radiation is therefore transmitted through a barium-filled region, so its detector signal differs more strongly from that of nearby tissue and the physical image contrast increases. After detection, a computer can map the recorded signal range onto a wider range of displayed pixel brightnesses. This makes an existing signal difference easier to see, but processing cannot recreate spatial information when the original detector signals contain no usable difference or are masked by noise. | 5 |
| 03.1 |
| Each absorbed X-ray produces charge carriers. The pixel signals are therefore and charge carriers. Both are below the saturation limit, and their ratio is , so a physical signal difference has been recorded. Digital processing can map that difference onto a wider displayed brightness range, but it cannot recreate information that the detector failed to distinguish because of saturation, noise or loss of spatial detail. | 5 |
| 04.1 |
| The time between frames must be at most , so the minimum frame rate is . In this records frames and gives relative dose units, or units to two significant figures. At the relative dose is units. The reduction is , or (accept ). Barium attenuates X-rays more strongly than neighbouring soft tissue, increasing the transmitted-intensity difference and making the marker visible. Below it can move more than between frames, so clinically relevant motion could be missed even though dose would fall further. | 6 |
| 05.1 |
| Each incident X-ray produces useful film exposures when the screen is used. The required incident number is therefore , or to two significant figures. Relative to the incident X-rays needed without the screen, the reduction factor is , or . The percentage reduction is , giving . Because dose is proportional to incident X-ray number, the patient dose falls by the same factor and percentage. | 4 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 | Use , which is to two significant figures. | 2 | |
| 02.1 | Use , which is to two significant figures. | 2 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 | Use . The exponent is . Therefore . | 3 | |
| 02.1 | Use . Taking natural logarithms gives , so the required thickness is to two significant figures. | 3 | |
| 03.1 | The attenuation exponents add: . Hence , so the absorbed fraction is . Therefore of the incident intensity is absorbed to two significant figures. | 3 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Attenuation exponents add for successive materials. For the bone path, , so . For soft tissue only, , so . Using unrounded values, the ratio is to two significant figures. | 5 |
| 02.1 | Convert the thickness: . From , . Since , , which is to two significant figures. | 4 | |
| 03.1 |
| Since , the exponent can be written as , where is the mass per unit area. For A, , or . For B, , or . The factor is , which is to two significant figures. | 4 |
| 04.1 |
| Dividing the two forms of eliminates : . Hence , giving . Using the first reading, , or . Finally, , which is to two significant figures. | 5 |
| 05.1 |
| The two paths differ only where of surrounding tissue is replaced by lesion. Thus . At L this is , giving to two significant figures; at H it is , giving . The all-surrounding transmissions are , or , and , or . Energy H is the only one above the transmission floor. Its smaller attenuation-coefficient difference gives weaker lesion contrast, so the increased transmission is obtained at the cost of a smaller detector-signal ratio. | 4 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Movement changes the beam direction, providing the many projections that the computer needs to reconstruct a cross-section. | 1 |
| 02.1 |
| The detectors convert the transmitted beam measurements into signals for computer reconstruction. | 1 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| The moving tube directs a narrow X-ray beam through the selected section from a succession of angles. The detector array measures the transmitted intensity for many paths. A computer processes this set of projections to calculate the attenuation distribution and display a cross-sectional image. | 3 |
| 02.1 |
| Restricting the beam width limits the region contributing to each detector reading, helping the reconstruction distinguish nearby structures. A monochromatic beam has a single photon energy and therefore a consistent attenuation coefficient for a given material. Variations in measured transmission can then be assigned more reliably to the tissues along each narrow path. | 3 |
| 03.1 |
| A single projection records only the total attenuation along each path through the patient. Different arrangements of structures along one path can therefore give the same reading and remain superposed. Rotating the source makes the X-rays cross each region along different paths, and combining these projections lets the reconstruction assign attenuation to separate positions within the slice. | 3 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| CT reconstructs slices, so ribs at different depths do not remain superimposed on the lesion and its three-dimensional location is clearer. That extra information can justify CT's higher ionising dose, cost and lower availability for the lesion. A plain radiograph is quick, cheap, widely available and gives a lower dose. Because a radiopaque tube already has strong contrast and only its overall position must be checked, the projection image is sufficient. Therefore use CT for the obscured lesion and a plain X-ray for the tube check. | 6 |
| 02.1 |
| Each angular projection requires X-rays to pass through the patient, so collecting fewer projections can lower the total ionising exposure. The computer then has less directional information, so fine detail and image resolution may deteriorate. A large lesion with clear boundaries may still be monitored adequately, but the protocol must retain enough resolution to detect a clinically important change. Its suitability depends on whether that limitation is outweighed by the safety advantage of a smaller accumulated dose from repeated scans. | 4 |
| 03.1 |
| CT reconstructs slices and can resolve fine detail without superposition, so it would provide more anatomical information. It is more expensive and normally exposes the patient to a greater X-ray dose than a plain radiograph. A simple forearm fracture has strong bone contrast, and the clinical target is only its overall alignment, so the extra CT resolution is unlikely to justify its added cost and dose across three examinations. Recommend plain radiographs unless they fail to show a feature needed for treatment. | 5 |
| 04.1 |
| A plain radiograph records attenuation along each complete X-ray path, so structures at different depths are superposed. Repeating the same type of projection may therefore leave the small lesion hidden even though the examination is relatively inexpensive and uses a lower dose. CT combines projections from many angles to reconstruct slices, removing superposition and locating the lesion with better spatial detail. This costs more and normally exposes the patient to a greater ionising dose. In this case the extra information has a stated surgical purpose and was not available from the radiograph, so the benefit of CT can justify its added cost and radiation risk. | 5 |
| 05.1 |
| The original number of contiguous slices is . Halving the slice thickness gives slices, which is twice the original number. Total dose is the dose per slice multiplied by the number of slices. Since the dose per slice is unchanged while the slice count doubles, the total patient dose also doubles. | 4 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| The required property is penetration: enough gamma photons must escape the tissue to reach the detector outside the patient. | 1 |
| 02.1 |
| The thyroid takes up iodine to make thyroid hormones, so labelled iodine becomes concentrated there. | 1 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Award one linked justification for each stated property. A half-life provides usable activity during the scan but causes activity to fall soon afterwards. The gamma radiation is penetrating enough for external detection while avoiding an unnecessarily high photon energy. Labelling a compound with organ affinity makes the recorded count distribution show that organ's function or structure. | 3 |
| 02.1 |
| Technetium-99m has a short half-life, so a transported stock would lose activity rapidly. The longer-lived molybdenum-99 parent remains in the generator and decays to technetium-99m. Staff can extract fresh daughter isotope when required, providing a practical local supply while retaining the dose advantage of the short-lived tracer. | 3 |
| 03.1 |
| Higher-energy gamma radiation is more penetrating. Some iodine-131 photons can cross lead between adjacent collimator holes instead of following the selected direction, so the recorded event may be assigned to an incorrect line and the image is less sharply resolved. They are also less likely to deposit all their energy in a finite crystal. Technetium-99m's lower-energy photons are therefore more readily stopped by the collimator where required and absorbed by the detector. | 3 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| The glucose analogue is carried into tissues, so regions with greater uptake contain more tracer. A radionuclide in the tracer emits a positron. After losing energy, the positron annihilates with an electron and produces two gamma photons travelling in approximately opposite directions. Detectors on opposite sides register near-simultaneous photons as a coincidence, locating the event somewhere along the joining line. A computer combines many lines of response to reconstruct the spatial distribution of uptake. | 5 |
| 02.1 |
| The observation lasts five technetium-99m half-lives, leaving only of its initial activity. It lasts indium-111 half-lives, leaving of the initial activity. Both isotopes emit penetrating gamma radiation and both can be attached to the required cells, but indium-111 maintains a detectable distribution for the full observation. Its longer effective residence in the patient prolongs irradiation, so the extra dose is justified only because the scan requires extended tracking. | 5 |
| 03.1 |
| Electron-positron annihilation produces two photons moving in nearly opposite directions. Detecting them within a coincidence window identifies the joining line of response. Without a measured arrival-time difference, no position along that line is selected, so a midpoint assignment invents information. The computer accumulates many lines; the region through which many of them pass is displayed as high uptake. Reducing detector-element width narrows the range of possible lines and improves spatial resolution, while the smaller collecting area lowers the number of accepted pairs per second. | 5 |
| 04.1 |
| For , , giving to two significant figures. For , , or . The reduction is , or . A smaller coincidence interval makes two unrelated photons less likely to be paired, reducing false lines of response. If the interval becomes narrower than the spread in arrival-time measurements, some true opposite-photon pairs will fall outside it and detection efficiency will decrease. | 5 |
| 05.1 |
| For X, the activity concentration in the organ is proportional to per kilogram, while the background concentration is per kilogram. Equal viewed masses therefore give ratio , or to two significant figures. For Y, the ratio is , or . Because photon energy, physical half-life, total activity and detector response are stated to be equal, the larger count ratio directly represents better contrast between the target and background. X is preferable because its labelled compound has greater affinity for the organ; its penetrating gamma photons can then carry that more concentrated distribution to the external detector. | 6 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| The definition must identify removal by biological processes and a reduction to one half; radioactive decay is excluded from this definition. | 2 |
| 02.1 |
| Physical half-life describes radioactive decay of the radionuclide itself. The definition requires a fall to one half and is independent of biological removal. | 2 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 | Use . Therefore , which is to two significant figures. | 3 | |
| 02.1 |
| The reciprocal relation adds the two positive removal rates: . Their sum is greater than either separate rate, so its reciprocal must be smaller than both and . The claimed exceeds and is impossible. | 3 |
| 03.1 | With both processes acting, the remaining fraction is . With radioactive decay alone it would be . The required fraction is , so remains relative to the physical-decay-only amount. | 3 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Rearrange the reciprocal relation: , so . The elapsed time is effective half-lives. The remaining activity fraction is therefore , giving . | 5 |
| 02.1 | First find the effective half-life: , so . Radioactive decrease gives . Taking logarithms, , which is to two significant figures. | 5 | |
| 03.1 |
| For physical decay, , so . The biological process supplies the additional factor in , so . Finally, , giving to two significant figures. | 4 |
| 04.1 |
| At , the three contributions are , and . Their sum is , giving to two significant figures. If is measured from the third administration, . Therefore , so the total falls below the threshold after to two significant figures. The effective half-life is used because both physical decay and biological removal reduce the activity in the patient. | 5 |
| 05.1 |
| For A, , so . For B, , so . Equality requires . Hence and . Substitution without rounding gives , or . Although B starts higher, its shorter biological and effective half-lives make its activity decrease more rapidly. | 6 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| The collimator selects photon directions before the scintillator; it does not focus gamma rays with refraction. | 1 |
| 02.1 |
| A gamma interaction in the crystal produces a flash of visible photons for the photomultiplier array. | 1 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| The gamma photon deposits energy in the scintillation crystal, which emits visible photons. Light reaching a photomultiplier photocathode releases electrons by the photoelectric effect. These electrons accelerate to successive dynodes and cause secondary emission at each stage, so a much larger electron pulse is collected at the anode. | 3 |
| 02.1 |
| The summed photomultiplier pulse is related to the energy deposited in the scintillator. Gamma photons scattered in the patient generally reach the camera with reduced energy and altered direction. An energy window rejects their lower pulse heights, preventing those events from being assigned to misleading positions and improving image contrast or resolution. | 3 |
| 03.1 |
| The larger share of event P's light reaches the first photomultiplier, so P is nearer that tube. Equal signals for Q place it midway between the two tubes. Position is inferred from the relative pulse sizes, while deposited energy is related to their sum; both events have total pulse size units and therefore correspond to the same deposited energy. | 3 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| The lead collimator accepts only a restricted range of directions, linking a detected photon to a line from the patient. A scintillation spreads light to several photomultipliers, and the relative sizes of their pulses allow the computer to estimate the interaction position. The total pulse height is related to deposited photon energy, so an energy window rejects lower-energy scattered photons that would be placed incorrectly. The computer accumulates the accepted positions to form the count image. Smaller collimator apertures sharpen directional selection and improve resolution, but fewer photons pass, so sensitivity falls. | 5 |
| 02.1 |
| After ten dynodes the electron number is . The collected charge is , which is to two significant figures. Each later dynode is more positive so that electrons accelerate towards it and arrive with enough kinetic energy to release several secondary electrons, sustaining the multiplication. | 4 |
| 03.1 |
| Halving the hole diameter reduces the resolution distance from to . Since , the two regions can be distinguished. The original count is . The new rate is , so the required time is . Narrower holes improve directional selection but reduce sensitivity, hence the longer acquisition. | 5 |
| 04.1 |
| Initially the excess is events and its uncertainty is , so the significance is , or . Treat the target count as background events plus an event excess. After narrowing the window, the background in each region is , while the retained excess is . The target and background counts are therefore and . The new significance is , or to two significant figures. Pulse-height selection has improved the evidence for localisation because lower-energy scattered events form much of the rejected background, despite the reduced total count. | 4 |
| 05.1 |
| Without the lead collimator, photons from a wide range of directions reach the scintillator. More events are detected, but their source directions are unknown, so accumulating them cannot produce a sharply located tracer distribution. Widening the pulse-height window also increases the recorded rate, but low-energy photons are commonly those scattered in the patient; scattering changes direction, so these events are placed misleadingly and degrade image contrast or resolution. If each absorbed gamma photon produces fewer visible photons, fewer photoelectrons are released at the photocathodes and the resulting anode pulse is smaller. Some genuine events may then fall below the counting threshold, while the greater relative effect of pulse fluctuations makes it harder to separate full-energy events from scattered events. The change therefore reduces useful sensitivity and weakens energy selection rather than improving the image. | 6 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| The required link is high photon energy to penetration to the depth of the tumour. | 1 |
| 02.1 |
| The shielding attenuates radiation outside the planned target region, reducing the dose received by healthy cells. | 1 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Aim each beam so that it passes through the tumour. At the tumour the energy deposited by all beam directions adds to the therapeutic dose. Away from the tumour, the entry and exit paths are spread through different healthy regions, so any one region receives only part of the total dose. | 3 |
| 02.1 |
| Imaging provides the positions and boundaries of the target and nearby sensitive structures. A computer uses this spatial information to choose beam directions and collimated field shapes. The beams are arranged to combine at the tumour while reducing the number or intensity of paths through healthy cells. | 3 |
| 03.1 |
| The unshaped field includes healthy-tissue cells, giving area . The matched field includes healthy-tissue cells, giving . The decrease is , or , about . Matching the field more closely to the tumour removes healthy tissue from the beam cross-section while retaining tumour coverage, so fewer healthy cells receive ionising radiation. | 4 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Since , the energy absorbed per session is . Over sessions, . Healthy-tissue exposure can be limited by using several beam directions that overlap at the tumour but spread entry dose, and by collimating or shielding the beam to exclude other tissue. Accurate imaging and computer planning improve targeting so that less healthy tissue lies within the treatment fields. | 5 |
| 02.1 |
| Each equal beam contributes of the tumour's total absorbed energy per kilogram. The healthy region receives of the tumour total because only one beam crosses it. A single beam would contribute the whole tumour total, so the same healthy region would receive . The four-direction plan therefore reduces this healthy-tissue exposure by a factor of . | 5 |
| 03.1 |
| For plan A, the organ dose is , or of the tumour dose. For plan B, the organ dose is , or . Both plans preserve the prescribed tumour dose, but plan B reduces the organ dose by percentage points and is therefore safer for that organ. | 5 |
| 04.1 |
| A deep target requires penetrating radiation so that useful energy reaches the tumour instead of being absorbed mainly near the surface. High-energy X-rays are ionising: the ionisation can damage tumour cells and prevent them from dividing successfully. Nothing in the interaction identifies a cell as malignant, so the same radiation can ionise and damage healthy cells on the beam path or beyond the tumour. Treatment therefore cannot make healthy-cell exposure exactly zero. Its use depends on delivering enough energy to control the tumour while keeping the unavoidable healthy-tissue risk as small as practicable. | 5 |
| 05.1 |
| The energy transferred per ionisation is . The number produced is , or to two significant figures. Ionisation causes cell damage, so the tumour cell may be unable to divide successfully. Ionisation is not selective: the same absorbed energy can damage a healthy cell. Accurate treatment planning is therefore required to concentrate the absorbed energy in the target and limit healthy-cell exposure. | 5 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Link the limited penetration of beta particles to a high local dose and reduced dose to distant healthy tissue. | 1 |
| 02.1 |
| Several source positions can match the tumour shape and reduce regions that are too far from a short-range beta source. | 1 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| The source is positioned within or adjacent to the target, avoiding a long entry path. Beta particles are strongly ionising over a limited range, so their energy is deposited close to the implant. High-energy X-rays are penetrating and an external beam necessarily irradiates some healthy tissue along its entry and exit paths. | 3 |
| 02.1 |
| Once the planned dose has been delivered, leaving the beta source in place would continue to ionise tissue. Removing it ends the exposure and limits damage to healthy cells near the tumour. | 2 |
| 03.1 |
| For A, . For B, , so the emitted particle totals are equal. Cells away lie beyond the stated range, showing why source positions must be distributed through a larger tumour. | 3 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| The elapsed number of half-lives is . Hence . Placing the source within or close to the tumour avoids irradiating a long path of healthy tissue before the radiation reaches its target. Beta particles have a short range in tissue, so most ionisation occurs near the implant and relatively little dose reaches distant cells. The source can also be removed when the planned treatment time is complete. | 5 |
| 02.1 | The number of decays is . The mean energy per decay is . The tumour absorbs . Hence , which is to two significant figures. | 5 | |
| 03.1 |
| The required absorbed energy is . The required number of decays is . The decay constant is and . From , to two significant figures. Multiplying by would treat the activity as constant at its initial value, although radioactive decay makes it decrease throughout the treatment. | 4 |
| 04.1 |
| Each implant emits . With two correct sources and one misplaced source, the tumour absorbs , so . Healthy tissue absorbs , giving . If all three are correctly positioned, the tumour dose is and the healthy-tissue dose is . Accurate placement both increases the intended local tumour dose and reduces irradiation of nearby healthy tissue. | 6 |
| 05.1 |
| The furthest points from the centre line are away. For a corner to lie within the beta range, the greatest allowed displacement along the centre line is . One source can therefore cover a length of the tumour, so an length requires at least sources. Positions , and make the end corners and the midpoints between sources exactly away or less. Equal activities give per source. Distributing short-range beta sources prevents distant parts of the tumour being underdosed while limiting irradiation beyond its boundary. | 6 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Ultrasound is portable, produces real-time images and does not use ionising radiation, so it meets all three requirements. | 1 |
| 02.1 |
| A plain X-ray image is quick and widely available, and bone gives strong X-ray contrast, so it suits the stated purpose despite the small ionising dose. | 1 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| MR distinguishes soft tissues well, so tumour boundaries can have high contrast. It uses magnetic fields and RF signals rather than ionising radiation, which is advantageous for repeated imaging. CT is often quicker and more readily available, but each scan contributes an X-ray dose; the justified preference is therefore MR unless speed or access is the overriding constraint. | 3 |
| 02.1 |
| MR distinguishes soft tissues with good spatial detail, so it can show the position and extent of a deep ligament injury. It does not use ionising radiation, but the scanner is expensive, less available and takes longer. Ultrasound is also non-ionising and is convenient for a portable, real-time examination. Its image quality depends more strongly on the operator and acoustic access, and its resolution of a deep ligament is generally lower than MR. | 3 |
| 03.1 |
| The tracer is chosen for its biological behaviour, and emitted gamma photons allow its changing distribution to be detected outside the body. Reduced uptake or clearance changes the recorded count pattern, providing functional information. CT instead maps the spatial distribution of X-ray attenuation, giving high-resolution structural information but not necessarily detecting an early functional change. | 3 |
| Question | Answers | Additional comments/Guidelines | Mark |
|---|---|---|---|
| 01.1 |
| Ultrasound is portable or widely accessible, quick and non-ionising, so it is suitable as an initial investigation, although its spatial resolution and image quality are lower and depend on the acoustic path. CT gives high-resolution cross-sectional localisation and strong contrast for dense stone material without superposition, but it uses a comparatively large ionising dose and is more costly. MR provides strong soft-tissue contrast without ionising radiation, yet calcified stone material is poorly shown, scans take longer and access is more limited. Therefore choose ultrasound first when it can answer the question, and use CT when the clinical need for precise localisation outweighs the radiation risk; MR is not the best choice for this target. | 6 |
| 02.1 |
| A PET tracer accumulates according to physiological activity, so unusually high uptake can provide direct evidence of active recurrence. PET requires an internally administered radioactive tracer, gives an ionising dose and generally has lower spatial resolution. MR provides strong soft-tissue contrast and detailed anatomical position without ionising radiation, making repeated structural imaging safer, but anatomy alone may not establish whether tissue is metabolically active. Recommend PET as the main test for the stated question of active recurrence, with MR used when precise anatomical localisation is also required. | 5 |
| 03.1 |
| PET is required for the activity map and resolves , smaller than the lesion. Pairing it with ultrasound gives total time and relative dose of the limit; ultrasound's structural resolution is also sufficient. MR plus PET would take and fail the time limit. CT alone gives times the limit, and CT plus PET gives times the limit. Ultrasound with PET is therefore the only pair satisfying information, resolution, time and safety constraints. | 5 |
| 04.1 |
| Avoiding ionising radiation is a genuine safety advantage, but a technique is useful only if it can obtain the required information. Ultrasound waves do not pass effectively through the skull, so ultrasound cannot replace CT for this purpose despite its portability and absence of ionising dose. MR can provide high-resolution soft-tissue information without ionising radiation, but the examination usually takes longer and may be impractical for some patients, which matters when a bleed must be found quickly. CT exposes the patient to ionising X-rays, yet it rapidly reconstructs cross-sectional images and avoids the superposition present in a plain radiograph. The decision therefore rests on resolution, diagnostic usefulness, convenience and safety. Cost may affect availability, but it is not needed to justify the clinical choice and carries no independent marking point here. | 6 |
| 05.1 |
| CT stores a spatial map reconstructed from X-ray attenuation measurements. Display processing can rescale contrast or emphasise boundaries already present, but attenuation alone does not record tracer uptake or metabolism, so it cannot be converted into a functional map. One completed CT acquisition also contains no sequence of later times, so software cannot make genuine live motion from the single dataset. The exposure occurred when the X-rays crossed the patient and cannot be removed afterwards. PET would add physiological information because a radioactive tracer accumulates according to activity, although it adds ionising dose and generally has lower spatial resolution. Ultrasound would add a real-time, non-ionising view of motion and is portable, but its usefulness depends on operator skill and an acoustic path not blocked by bone or gas. These methods complement rather than retrospectively transform the CT data. | 6 |